Heat of Combustion Tester for Building Material

China Heat of Combustion Tester for Building Material, Find details about China En ISO 1716 Bomb Calorimeter, Heat of Combustion Tester from Heat of Combustion Tester for Building Material

Model NO.
GD-ISO1716
Heat Repeatability Error
≤0.2%
Heat Capacity of The Instrument
About 10000j / K
Ambient Temperature
(10 ~ 35) ºC, Humidity ≤80%
Maximum Power
0.5kw
Oxygen Bomb Pressure Range
0.5 ~ 3.5MPa, No Leakage. The Bomb Pressure 21MPa
Bomb Capacity
(300 ± 50) Ml
Warranty
12 Months
Trademark
GOLD
Transport Package
Plywood Case
Origin
Chongqing China(Mainland)
HS Code
9031809090
Model NO.
GD-ISO1716
Heat Repeatability Error
≤0.2%
Heat Capacity of The Instrument
About 10000j / K
Ambient Temperature
(10 ~ 35) ºC, Humidity ≤80%
Maximum Power
0.5kw
Oxygen Bomb Pressure Range
0.5 ~ 3.5MPa, No Leakage. The Bomb Pressure 21MPa
Bomb Capacity
(300 ± 50) Ml
Warranty
12 Months
Trademark
GOLD
Transport Package
Plywood Case
Origin
Chongqing China(Mainland)
HS Code
9031809090
Heat of Combustion Tester for Building Material

Introduction: 

Heat of combustion of materials is important characterize parameters to calculated building materials release heat and fire basic data.The bomb calorimeter is designed according to ISO 1716-2002.

 

Standard: 

  • ISO 1716-2002

  •GB/T 14402-2007

  •GB/T8624-2012

 

Feature: 

1, The apparatus including calorimeter apparatus, high-pressure oxygen bomb, cold water installations, computer and control system (which can record and display the curve), LCD screen and software;

Crucible made of stainless steel;

 2, Box paint treatment; all internal parts are made of stainless steel, corrosion-resistant, easy to clean;

 3, Ignition wire break is the detection and automatic ignition.

 4, The test procedure automatically turn off.

5, Test abnormalities has a sound alarm.

6, Computer control system:

7, Control box with USB / RS232 serial port with an external brand desktop or laptop operating control; High degree of automation, test procedure safe and reliable, easy to operate;

 

Specification: 

 Bomb capacity

  (300 ± 50) mL.

 Oxygen bomb pressure range

  0.5 ~ 3.5MPa, no leakage. The bomb pressure 21Mpa.

Timer

  accuracy of 0.5s;

Maximum power

  0.5KW;

Ambient temperature

  (10 ~ 35) ºC, humidity ≤80%;

 Heat capacity of the instrument

  about 10000J / K;

Heat repeatability error

  ≤0.2%;

Temperature range

  (4.5-50) ºC;

Temperature detection

  0.0001 ºC resolution, using the PT100 sensor;

Inner cylinder volume

 2000ml; Oxygen bomb Sealing

Motor power

  3W; stirring speed

Power supply voltage

  AC220V ± 10% 50Hz


Heat of Combustion Tester for Building Material

Heat of Combustion Tester for Building Material