China Non-Combustibility Test Apparatus for Building Materials ISO 1182 BS 476-4 BS 476-11, Find details about China Non-Combustibility Test, Non-Combustibility Test Apparatus from Non-Combustibility Test Apparatus for Building Materials ISO 1182 BS 476-4 BS 476-11
Introduction:
Non-Combustibility Tester measures incombustibility and combustibility of uniform building material and non-uniform building material of actual element.
The evaluations (results) are rising temperature in the furnace, basic mass loss and ignition of a specimen.
It fully satisfies KS F 2271, KS F ISO 1182 and ISO 1182 as international standards in the same time.
This equipment uses a module type temperature controller to be stable over 60 minutes in the furnace condition of 750±5ºC and meets precise test result.
And another advantage is the automatic up-down specimen movement device for accurate results as well as a user-friendly process. Thus, it doesn't matter who is test-users.
RS 232C communication and Data Acquisition Board is able to be accurate Interface of test values, so user can test the whole function and test by PC easily and automatically.
Application:
* Building Materials
* Rail
* IMO
Standards:
1. ISO 1182: Reaction to fire test for building material---Non-combustibility test
2. BS 476-4 : Fire Tests on Building Materials and Structures.. Part 4: Non-Combustibility Test for Materials
3. BS 476-11: Fire tests on building materials and structures - Part 11: Method for assessing the heat emission from building materials.
4. KS F ISO 1182: Method of a non-combustibility test of building products
5. KS F 2271: a Testing method for incombustibility of internal material and element of building
Feature:
1. Single Zone Furnace operable at up to 900 ºC with the NIKROTHAL Ribbon Heater
2. 750±5ºC temperature auto control with TPR
3. Furnace facilitating maintenance and replacement
4. Up/down automatic moving system for correct positioning of test sample and user's convenience
5. Mirror to observe the ignition quality of test sample
6. RS 232C communication and DAQ Board-adopted interface
7. Auto testing by DAQ Board and PC
8. Alarm temperature-based over current shut-off system for user's safety
9. Software (Non-Comb 2005) to print out reports based on exact data result and each standard
Test Result:
* Mass Loss(g) and Rate(%)
* Ignition Time(s) and Sustained Flaming(s)
* Initial & Maximum Furnace Temperature(ºC)
* Final Furnace Temperature(ºC)
* Maximum Specimen Center Temperature(ºC)
* Final Specimen Center Temperature(ºC)
* Furnace Temperature Rise(ºC)
* Specimen Center & Surface Temperature Rise(ºC)
* Differential Test Result by Standards
(ISO 1182:1990, KS F ISO 1182: 2004, IMO FTPC Pt.1: 1998, KS F 2271: 1998)
Specifications:
Size | 400mm(w) × 40mm0(D) × 1,270mm(H) |
Power | 220 V AC, 60 Hz, 10 A |
Weight | about 40kg |
control unit | 530mm(w) × 270mm(D) × 280mm(H) |
Manual | supplied |
Utilities | Personal Computer and PC Printer |