China Ce Marked Veterinary Anesthesia Machine with Ventilator Anaesthesia Machine, Find details about China Veterinary Anesthesia Machine, Animal Anesthesia Machine from Ce Marked Veterinary Anesthesia Machine with Ventilator Anaesthesia Machine
Ventilation modes: | IPPV;A/C;SIMV; |
Ventilator parameter ranges | |
Flow meter | O2(0-5L/min) |
Rapid oxygen supply | 35L/min-75L/min |
Tidal volume(Vt) | 0, 20 mL ~ 1500 mL |
Frequence(Freq) | 1bpm ~ 100 bpm |
I/E | 2:1~1:6 |
Pressure triggering sensitivity(PTr) | -20 cmH2O ~ 0 cmH2O(Based on PEEP) |
Flow trigger sensitivity (FTr) | 0.5 L/min ~ 30 L/min |
SIGH | 0(off) 1/100 ~ 5/100 |
Apnea Ventilation | OFF,5 s ~ 60 s |
Pressure Limit | 20 cmH2O ~ 100 cmH2O |
Monitored Parameters | |
Frequence(Freq) | 0 /min ~ 100 /min |
Tidal volume(Vt) | 0 mL ~ 2000 mL |
MV | 0 L/min ~ 100 L/min |
Graphical display: | |
P-T(pressure - time) | |
F-T(flow - time) | |
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | <11.3±0.3V |
No tidal volume | ≤5Ml within 6s |
High oxygen concentration alarm Lowoxygen concentration alarm | 19%-100% 18%-99% |
High Airway pressure alarm Low Airway pressure alarm High Minute Volume alarm Low Minute Volume alarm Continuous Pressure alarm | 20cmH2O-100cmH2O 0cmH2O-20cmH2O Adult(5L/min-20L/min) Paed(1L/min-15L/min 0-10L/min) (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
The maximum limited pressure | <12.5 kPa |
Fan error Oxygen deficit | Show on screen Show on screen |
Working conditions | |
Gas source | O2 |
Pressure | 280kPa-600kPa |
Voltage | 100-240V |
Power frequency | 50/60Hz |
Input power | 60VA |