Flu a B Diagnostics Test Influenza (Grippe) a+B Rapid Test Kit

China Flu a B Diagnostics Test Influenza (Grippe) a+B Rapid Test Kit, Find details about China Flu a B Rapid Diagnostic Test Kit, Influenza a+B Rapid Test Kit from Flu a B Diagnostics Test Influenza (Grippe) a+B Rapid Test Kit

Model NO.
FLU0095
Storage
2′c-30′c
After-Sale Service
Return and Replacement
Method
Colloidal Gold Rapid Test
Trademark
Neo nostics
Transport Package
Carton
Specification
Colloidal Gold Method Flu Test
Origin
Jiangsu, China
Model NO.
FLU0095
Storage
2′c-30′c
After-Sale Service
Return and Replacement
Method
Colloidal Gold Rapid Test
Trademark
Neo nostics
Transport Package
Carton
Specification
Colloidal Gold Method Flu Test
Origin
Jiangsu, China
Flu a B Diagnostics Test Influenza (Grippe) a+B Rapid Test Kit
Influenza A + B Test Kit
(Swab/Nasal Aspirate)
Material
Provided
Test device
Buffer
Desiccant pouch
Extraction tube
Swab
Not provided
Timer
 
The Influenza A/B test is a rapid chromatographic immunoassay for the qualitative detection of influenza A and B antigens in nasal swab  or nasal aspirate specimens. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. A rapid test for the qualitative detection of Influenza A and Influenza B virus in nasal swab or nasal aspirate specimens.
 



The kit used for in vitro qualitative test of Influenza A/B and nasal antigen samples.
Flu a B Diagnostics Test Influenza (Grippe) a+B Rapid Test Kit
----------------------------------------------------------------------------------------------------------------------------------------------------------
Method of use
 
1. Remove the test kit from the sealed foil pouch and use it as soon as possible. Best results will be obtained if the assay is performed immediately after opening the foil pouch.
2. Place the Extraction Tube in the workstation. Hold the extraction reagent bottle upside down vertically. Squeeze the bottle and let the solution drop into the extraction tube freely without touching the edge of the tube. Add 10 drops of solution (Approx. 400ul) to the Extraction Tube.
3. Place the swab specimen in the Extraction Tube. Rotate the swab for approximately 10 seconds while pressing the head against the inside of the tube to release the antigen in the swab.
4. Remove the swab while squeezing the swab head against the inside of the Extraction Tube as you remove it to expel as much liquid as possible from the swab. Discard the swab in accordance with your biohazard waste disposal protocol.
5. Fit the dropper tip on top of the extraction tube. Place the cassette on a clean and level surface.
6. Add three drops of the solution (approx.120ul) to the sample well and then start the timer. Read the result at 15 minutes. Do not interpret the result after 20 minutes.
Flu a B Diagnostics Test Influenza (Grippe) a+B Rapid Test Kit
----------------------------------------------------------------------------------------------------------------------------------------------------------
Production process
Flu a B Diagnostics Test Influenza (Grippe) a+B Rapid Test Kit
 
----------------------------------------------------------------------------------------------------------------------------------------------------------
Certification
Flu a B Diagnostics Test Influenza (Grippe) a+B Rapid Test Kit
----------------------------------------------------------------------------------------------------------------------------------------------------------
FAQ

Q1.What's the method you use to do the test?

A: Colloidal Gold Method.

Q2: What's ur MOQ ?
A: Always, for strip format, the MOQ is 5000pcs. While for cassette format, MOQ is 2000pcs, same as midstream.  

Q3: What's your packing of the tests?
A: If you choose the low cost, we will suggest you the bulk packing to go. Which means 1 test in a single pouch, then 100/25 pouches in a plastic bag, 50 bags in a carton. If you need to sell in pharmacy/chain shop or some countries which have strict requirements about the packing, you can choose to pack in a box, like 30 pcs in a box, or even 1 test in a single box.

Q4: What's the main differences when comparing products from different suppliers?
A: You can compare the mainly features of the test: Accuracy, Sensitivity, Specificity, and then the Price.

Q5: Some tests can be tested with serum/plasma or with whole blood, so what's the difference?
A: Well, if the specimen is whole blood, people will need buffer, lancet and alcohol swab to help the test. If they buy all the accessories, they can test very easily at home. But if the specimen is serum/plasma, it will need centrifuge to separate serum from whole blood first. This way, it's better to use at lab or hospital. And in some words, serum/plasma test will always give a more accurate results.

Q6: How can i distinguish a good test kit?
A: You can judge from the 4 facts: 
1. Technical data: Such as the accuracy, specificity and sensitivity.
2. Pouch sealing: Tight enough. If the foil pouch is not sealed well, the humidity in circustance will destroy the reactivity of antibodies labeled on NC membrane. Shelf life will be shorten down.
3. Flow speed: The shorter the better??  NO!!   The reaction of antibodies on NC membrane and antigens in specimen usually requires quite a while to work sufficiently. You can refer to the instruction for more accurate time.
4. Background: Good test usually gives clean background after running. If there are red smears in the reading window, it usually caused by bad colloidal gold technology or bad NC membrane. Sometimes, the defect caused false positive result in practice.