Biobase China Clinical Lab Using Microplate Auto Elisa Processor, Find details about China Fully Automatic Elisa Processors, Elisa Processor from Biobase China Clinical Lab Using Microplate Auto Elisa Processor
Features·
Model | BIOBASE 1000 | |
Sample Unit | Sample Racks | 80 sample positions(tubes 13mm) |
Pipetting | 10~1000 | |
Dispensing System | 1 aspirating-dispensing probe with X-Y-Z movement; | |
Reagent Unit | Reagent Racks | 8+1reagent racks, editable |
Pipetting Range | 10~1000, 1 stepping | |
Pipetting precision | (100vl) CV≤0.5% | |
DispensingTime | 5 minutesto whole96well microplates | |
Incubation Unit | Incubation | 2 independentIncubators |
Temperature Range | RT to 45 °C | |
Temperature Accuracy | ±0.5°C | |
Washing Unit | WashingProbes | Two line 8-nozzle manifolds |
Wash Containers | 2 wash buffers 2L, 1 distilled water, 1 waste water | |
Washing Residual | <2 | |
Reading Unit | ReadingChannels | 8 independent photometric channels,mono and dichromatic reading |
Absorbance | 0.0 ~3.0 OD | |
Spectral Range | 400-700nm | |
Optical Filters | 2standardfilters (450,630nm); 2 more on request (405nm/450nm/492nm/630nm) | |
Data Management | Software | Windows 7 |
LIS System | Bi-direction,support HL7 protocol | |
Results | Absorbance and results reviewable by software | |
Printer | External optional | |
Working Conditions | Power Supply | AC220V±10%,50/60Hz, 110V±10%,60Hz, 400W |
Temperature | 10~30ºC | |
Humidity | 30-80% | |
Size & Weight | External Size (W*D*H) | 879*660*764mm |
Net Weight | 112kg | |
Package Size (W*D*H) | 1020*810*1240mm | |
Gross Weight | 160kg |