Hbhkn-2000 Delivery Room Baby Neonatal Infant Radiant Warmer Incubator

China Hbhkn-2000 Delivery Room Baby Neonatal Infant Radiant Warmer Incubator, Find details about China Radiant Incubator, Warmer Incubator from Hbhkn-2000 Delivery Room Baby Neonatal Infant Radiant Warmer Incubator

Model NO.
HBHKN-2000
I.V. Pole
Yes
Temperature Sensor Display Range
5-65 Degree
Apgar Timer
Yes
Failure Alarm
Yes
Trademark
WANROOEMED
Transport Package
Export Carton Package
Specification
111cm× 62.5cm× 119.5cm
Origin
China
HS Code
94029000
Model NO.
HBHKN-2000
I.V. Pole
Yes
Temperature Sensor Display Range
5-65 Degree
Apgar Timer
Yes
Failure Alarm
Yes
Trademark
WANROOEMED
Transport Package
Export Carton Package
Specification
111cm× 62.5cm× 119.5cm
Origin
China
HS Code
94029000
HBHKN-2000 Delivery Room Baby Neonatal Infant Radiant Warmer Incubator
Hbhkn-2000 Delivery Room Baby Neonatal Infant Radiant Warmer Incubator

Brief Introduction:

Pre-warm mode, Manual mode and Baby mode;

Set temperature and skin temperature can be displayed separately;

Main body is made of Stainless steel;

Double sides phototherapy radiant box;

Acrylic glass bassinet;

APGAR timer;

Timer to count phototherapy time;

Failure alarm indication;

RS232 connector.

Standard configuration

It is mainly composed of the warmer module, controller, bassinet, phototherapy, and the stand and so on.

Specification

Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 850VA

Mode of control: Pre-warm mode, manual mode and baby mode are controlled by micro-computer

Baby mode temperature control range: 34.5-37.5°C

Skin temperature display range: 5°C -65°C

Accuracy of skin temperature sensor: ±0.2°C

Temperature uniformity of mattress≤2°C

Dominant wavelength of blue light: 400-550nm

Timing range of blue light: 0-9999hrs 59min

APGAR timer: audible and visible alarm when the device runs to 50" - 1', 4' 50" - 5', 9' 50" - 10' ;

Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, power failure alarm, setting alarm, checking alarm and so on.

Average value of general bilirubin radiation on the mattress effective range:

≥250uW/cm2 ( the above phototherapy equipment)

≥600uW/cm2(the nether phototherapy equipment)

Environmental temperature

Operating range: 18ºC-30ºC

Ambient air movement rate: ≤0.3m/s

Package

Package: each unit is packed to one carton

Size(Main body ): 94cm×57cm×108cm

Gross Weight:57kg



Details:
Hbhkn-2000 Delivery Room Baby Neonatal Infant Radiant Warmer Incubator