STANDARD FEATURES:
1. Servo Control Mode: air temperature, skin temperature, display humidity;
2. Parameter Displaying: set temperature, air temperature, skin temperature, internal humidity, heat power rate;
3. Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air & skin temperature override, fan failure, system failure;
4. >37ºCPreset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;
5. Hood with dual slope; 6 hand-ports;
6. RS-232 Output Connector (use for network system);
7. IV poles;
8. Protective guard rail;
9. Storage cabinets.
OPTIONS:
Height adjustable stand
Humidity control system
LCD display screen
Specifications of hood
Mattress to Hood: 450 mm
Observation side of hood: Two incline
Operate port: 6 (2 Iris ports)
Tubing grommets: 4
Front access panel: 826×310 mm
Mattress Tray Size: 630×360 mm
Technical Specifications:
Power requirement ≤400VA
Temperature control mode: Air and skin servo control
Air Temp Control Range: 20ºC~37ºC
Temperature override mode to: 20ºC~39ºC
Skin Temp Control Range: 30ºC~37ºC
Temperature override mode to: 30ºC~38ºC
Skin temp sensor accuracy: ≤0.3ºC
Humidity display range: 5%~99%
Temperature fluctuation: ≤0.5ºC
Mattress temperature uniformity: ≤0.8ºC
Temperature rise time: ≤45 min
Noise level within hood environment: ≤55dB
Alarms
Power failure alarm
Fan failure alarm
Sensor failure alarm
High air temperature alarm: +3.0ºC
Low air temperature alarm: -3.0ºC
High skin temperature alarm: +1.0ºC
Low skin temperature alarm: -1.0ºC
Over air temperature alarm: ≤38.0ºC; ≤40.0ºC
Over skin temperature alarm: ≤38.0ºC; ≤39.0ºC
System failure alarm
Alarm silence function
Operational condition
Environment temperature: +20ºC~+30ºC
Environment relative humidity: 5%~99%
Environment air velocity of flow: <0.3m/s
Others
Infant bed tilt adjusting ±10° step-less
Mattress tray Max. load 10kg
Transfusion pole Max. load 2kg
Cabinet ax. load 6kg