China Medical Device Hospital Portable Automatic Transport Neonatal Care Infant Incubator, Find details about China Infant Incubator, Baby Incubator from Medical Device Hospital Portable Automatic Transport Neonatal Care Infant Incubator
Features:
BT-100 Transport Incubator
Air mode and Baby mode controlled by micro-computer
Use AC Power and DC Power alternatively, DC12V or DC 24V Power on the ambulance
Set temperature, air temperature, Skin temperature and internal battery Power can be displayed by LED separately
Double wall hood with side door
The infant bed can be pulled out easily
Natural air flow humidity
The height of whole unit is adjustable
Oxygen cylinder and Oxygen supply system
LED observe light, adjustable brightness
Optional trolley:Ambulance type trolley
Brief introduction:
Air mode and Baby mode controlled by micro-computer;
Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance;
Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately;
Failure alarm indication;
Double wall Acrylic glass hood with side door so that the bassinet can be pulled out from it;
Natural air flow humidity;
The height of whole unit can be adjusted;
With Oxygen cylinder and Oxygen supply configuration;
Observation lamp.
Standard Configuration:
Main body (including the hood,controller,bassinet,internal battery,and observation lamp),oxygen supply system,skin temperature sensor,I.V.Pole,mattress,adjustable stand)
Optional Configuration:>37°C override temperature mode
Specification:
AC Power supply:AC220V-230V/50HZ
DC Power supply:DC12V/10A or 24V/6A
Input Power:400VA
Mode of control: Air mode and Baby mode controlled by micro-computer
Air temperature control range:25°C-37°C
(Override mode):37-38°C
Baby temperature control range: 34°C-37°C
(Override mode):37-37.5°C
Skin temperature sensor accuracy: ≤0.3°C
Temperature variability : ≤1.0°C
Temperature uniformity : ≤1.5°C(mattress is on the horizontal position)
Noise inside hood : ≤55dB(A) (Environment noise ≤45Db(A) )
Failure alarm :over-temperature alarm, temperature diviation alarm,sensor alarm,motor fan failed alarm,power failure alarm and so on.
Operating at temperature equilibrium :90 min (Set Temp 36°C, Ambient 15°C)
Environmental temperature:
Operating range: 10°C-30°C
(Set temperature should be higher 3°C than the environment temperature)
Ambient air movement rate : ≤1.0m/s
Factory