STANDARD COMPOSITION:
1. Servo Control Mode: air temp;
2. Instant Parameter Displaying: set temp, air temp, heat power rate;
3. Audible & Visual Alarms: power failure, sensor failure, higher/lower temp, air temperature override, fan failure;
4. Power failure reset memory and self-test function;
5. Hood with dual slope: 4hand-ports.
Specifications of hood
Mattress to Hood: 450 mm
Observation side of hood: Two incline
Operate port: 4 (2 Iris ports)
Tubing grommets: 2
Front access panel: 826×310 mm
Mattress Tray Size: 630×360 mm
Technical Specifications:
Power requirement ≤400VA
Temperature control mode: Air servo control
Air Temp Control Range: 20ºC~37ºC
Temperature fluctuation: ≤0.5ºC
Mattress temperature uniformity: ≤0.8ºC
Temperature rise time: ≤45 min
Noise level within hood environment: ≤55dB
Alarms
Power failure alarm
Fan failure alarm
Sensor failure alarm
High air temperature alarm: +3.0ºC
Low air temperature alarm: -3.0ºC
Over air temperature alarm: ≤38.0ºC
Operational condition
Environment temperature: +20ºC~+30ºC
Environment relative humidity: 5%~99%
Environment air velocity of flow: <0.3m/s
Others
Mattress tray Max. load 10kg
Transfusion pole Max. load 2kg