THR-PM2000 Portable Medical Fetus Monitor

 
Description
FHR(Fetal Heart Rate)

Method: Ultrasound Doppler

Measuring Range: 60-200bpm

Precision: ≤±2bpm

Ultrasound Frequency: 2.5 MHz

Ultrasound output: 5mW/cm2

Ultrasound Task Measurement:Impulse Wave

Alarm Indication range: upper limit:66-200bpm

Lower limit: 65-199bpm

TOCO (Uterine Contraction Pressure)

Method: Outside Measurement

Measuring Range: 0-100 unit

Non-linearity Precision: ≤±10%

Quickening Sign
Quickening Sign: Artificial and Automatic

SPO2

Sensor:Two-waveform LED(light emitting diode)

SPO2 measuring range:0-100%

Resolution:1%

Precision: ≤±2%(90%-100%);≤±4%(70%-89%);≤±6%(40%-69%)

Alarm Indication range:Upper limit 1-100%;Lower limit 0-99%

Range of pulse rate:20-250 bpm

Pulse alarm indication range:Upper limit 22-250 bpm;Lower limit 20-248 bpm

NIBP

Method: Intelligent Oscillation

Cuff inflating time: 30s (standard adult cuff 0-250 mmHg)

Deflating time: average90sec

Deflating time when measurement stop:2sec(standard adult cuff 250-150mmmHg)

Initial inflating pressure:180mmHg(adult); 120mmHg(infant)

Maximum allowed inflating pressure:300mmHg(40Kpa,over this value, safety

Value deflating automatically)

Measuring Mode:Manual measurement, adult mode

Measurement Intermission:0-240min adjustable

Measurement parameters:Systolic pressure,Diastolic pressure,Mean pressure

Alarm indication range:Upper limit: 22-250mmHg;Lower limit: 20-248mmHg

Alarm Indication

Alarm indication range of FHR, UCP.HR, NIBP SPO2 are settable.

Sound and light alarm are available.

 

Other Specification

Power supply:220V±22V,50Hz±1Hz

Power consumption: ≤110VA

Display mode: 12.1 inch TFT LCD display

Alarm mode: Sound and Light

Working condition: Ambient temperature: 5-40

Relative humidity: ≤80%

Atmospheric pressure: 86-106Kpa

Storing condition: Ambient temperature: -10-40

Relative humidity: ≤80%

Keep away from acrid gas

Good draught

Safety: Class I, Type BF