Model NO.
PA-700B
Simv Rate
1-20bpm
Oxygen Concentration
21-100%
Pressure Limit
1.0kpa~6.0kpa
OEM
Yes
MOQ
1
Kind
All
Group
≥5kg
LCD Display
TFT Display
Trademark
Perlong
Specification
0.55CBM 81KG
Origin
China
HS Code
9019200000
Model NO.
PA-700B
Simv Rate
1-20bpm
Oxygen Concentration
21-100%
Pressure Limit
1.0kpa~6.0kpa
OEM
Yes
MOQ
1
Kind
All
Group
≥5kg
LCD Display
TFT Display
Trademark
Perlong
Specification
0.55CBM 81KG
Origin
China
HS Code
9019200000
                           PA-700B Medical Ventilator

Specification
Model700B (Advanced Model)
Display10.4 inch high visibility color TFT display
Ventilation ModeA/C, IPPV, SIPPV, IMV, SIMV, Standby, SPONT MANUAL
Minute VolumeGreater than or equal to 18L
Back-up Power SupplyAt least 4 hours
Tidal VolumeAdjustable range: 50-1500ml, Display range: 0-2000ml
Respiratory Rate1-99bpm
Inspiratory/expiratory (I:E) Ratio4:1-1:4
PEEP0-10cmH2O 
SIMV Rate1-20bpm
Inspiratory Pressure Trigger.-10-10cmH2O 
Inspiratory PlateauAdjustable from 0-50% of inspiratory time
Adjustable Oxygen Concentration21-100%
Gas Supply Requirement280~600 kPa medical grade oxygen and compressed air supply
Pressure Limit1.0kPa~6.0kPa
Maximum Safe Operating Pressure≤6.0kPa
Tidal Volume AlarmUpper Limit: 10~2000ml, Lower Limit: 0~1800ml
Airway Pressure AlarmUpper Limit: 0.1~6kPa, Lower Limit: 0~5kPa
Oxygen Concentration AlarmUpper Limit: 22~100%, Lower Limit: 21~80%
 
 
Features
Operating ModesPneumatically driven & electronically controlled, closed, semi-closed and semi-open
Application1. emergency clinical resuscitation                               2. respiratory therapy
DisplayBuilt-in 10.4 inch high visibility color TFT display, able-simultaneously display Pressure-Time and Frequency-Time waveforms in real time
Sighinsert 1-8 sigh breaths in every 100 breaths, ventilation should be no less than 1.5 the tidal volume
Equipmentinfrared turbine flow sensor; precisely measures gas flows, stable and reliable
Power Failure Support FunctionsAutomatically converts-standby mode during AC power outages
Humidifier9 Level adjustable humidifier
Monitor ParametersTidal volume, respiration rate,                                       Airway pressure,                                                    
I:E ratio,                                                                   Inspiratory & expiratory status, inspiratory trigger and others
AlarmsAudible and/or visual alarms for ventilation volume, tidal volume, oxygen concentration, asphyxia, airway pressure, intubation disconnection, low oxygen, power failure and other alarms, also features a alarm history query function
Advanced Model Exclusive FeaturesEquipped with professional medical grade air compressor tailored for long clinical times and low noise requirements