China Biometer Multiple Functions Offline Verification Toolkit for Dissolution Tester, Find details about China Test Machine, Pharmaceutical Machinery from Biometer Multiple Functions Offline Verification Toolkit for Dissolution Tester
The special physical parameter correction tool for the dissolution tester is not only suitable for the daily periodic mechanical calibration of the dissolution tester, but also can be used to exclude specific physical parameters that may produce abnormal test results-it can check whether there are physical parameters that have changed or are out of range.
Physical parameters which can be measured include:
1.RPM
2.Rotating shaft swinging degree
3.Basket swinging degree
4.Rotating shaft perpendicularity
5.Dissolution cup perpendicularity
6.Dissolution cup centrality (upper position and lower position)
7.Basket/paddle bottom height
8.Levelness (dissolution tester, table)
9.Temperature of water bath tank and cups
10.Vibration degree (four parameters: accelerated speed, frequency, displacement and speed)
Verification Parameters | Technical Requirements |
Dissolution Tester Levelness | ≤0.5° |
Basket (paddle) Axis Perpendicularity | 90.0°±0.5° |
Dissolution Cup Perpendicularity | 90.0°±1° |
Coaxiality of the Dissolution Cup and the Basket (Paddle) | ≤2.0mm |
Swing of Basket (Paddle) Axis | ≤1.0mm |
Basket Swing | ≤1.0mm |
Depth of Basket (Paddle) | 25±2mm |
Rotating Speed of Basket (Paddle) | ±4% |
Dissolution Cup Temperature | 37°C ±0.5°C |
Coaxiality Measurement Dialgage | Indexing: 0.01mm, measuring range: (0-3)mm |
Swing Measurement Dialgage | Indexing: 0.01mm, measuring range: (0-3)mm |
Calliper | Indexing: 0.01mm, measuring range: (0-150)mm |
Thermometer | Resolution 0.01°C/0.1°C, 0.01°F/0.1°F Accuracy: ±(0.1%+0.2°C) |
Tachometer | Accuracy: 0.04%±2dgts±0.06RPM |
Inclinometer | Resolution 0.05° |