Model | BIOBASE 2000 | ||||
Sample Unit | Sample Racks | 552 sample positions(tubes 13mm) | |||
Pipetting | 8~1000μl | ||||
Dispensing system | 2 aspirating-dispensing probe( X-Y-Z movement) | ||||
Reagent Unit | Reagent Position | 23 reagent racks, editable | |||
Pipetting Range | 8~1000μl, 1μl stepping | ||||
Pipetting Precision | (100μl )CV≤0.5% | ||||
Washing Unit | Dispensing Time | 4 minutes to whole 96 to microplates | |||
Washing Probes | Two line 8-nozzle manifolds | ||||
Wash Containers | Cleaning fluid(15L),waste water(25L),Buffer 1(5L), Buffer 2(2L), Buffer 3(2L), Buffer 4(2L), with liquid level-sensing, liquid shortage and full alarm | ||||
Waste Containers | 25L with waste full sensor | ||||
Washing Residual | <2μl | ||||
Reading Unit | Reading Channels | 8 independent photometric channels, mono anddichromatic reading | |||
Absorbance | 0.0~3.0 OD | ||||
Spectral Range | 400-700nm | ||||
Optical Filters | 2standard filters (450,630nm); 2 more on request (405nm/450nm/492nm/630nm) | ||||
Reading Accuracy | ±1% from 0.0~1.5 OD, ±2% from 1.5 to 3.0 OD | ||||
Data Management | Software System | Windows 7 | |||
LIS System | Bi-direction,support HL7 protocol | ||||
Results | Absorbance and results reviewable by software | ||||
Printer | External optional | ||||
Working Conditions | Power Supply | AC220V±10%,50/60Hz, 110V±10%,60Hz, 400W | |||
Temperature | 10~30ºC | ||||
Humidity | 30-80% | ||||
Size & Weight | External Size (W*D*H) | 1235*710*880mm | |||
Net Weight | 146kg | Gross Weight | 160kg | ||
Package Size (W*D*H) | 1380*860*1435mm | ||||
Gross Weight | 220kg |