Oxygen Index Apparatus / Limiting Oxygen Index

China Oxygen Index Apparatus / Limiting Oxygen Index, Find details about China ISO4589, ASTM D 2863 from Oxygen Index Apparatus / Limiting Oxygen Index

Model NO.
FTech-ISO 4589
Control
Computer Control
Weight
100-500Kg
Power Source
AC220V
Oil Cylinder Position
Top
Standards
ASTM D 2863, ISO 4589
Warranty
1 Year
Dimension
435(W)*550(D)*670(H)mm
Trademark
TESTech
Transport Package
Plywood Case
Origin
China
HS Code
90318090
Model NO.
FTech-ISO 4589
Control
Computer Control
Weight
100-500Kg
Power Source
AC220V
Oil Cylinder Position
Top
Standards
ASTM D 2863, ISO 4589
Warranty
1 Year
Dimension
435(W)*550(D)*670(H)mm
Trademark
TESTech
Transport Package
Plywood Case
Origin
China
HS Code
90318090
Introduction:
TESTECH Limited Oxyen Index Tester for the determination under the test conditions specified in oxygen and nitrogen gas mixture just to maintain a minimum sample combustion oxygen concentration required. This tester equipped with the imported oxygen concentration sensor, to make sure the measurement accuracy and durability. For the assessment of the performance of a homogeneous combustion of solid materials, laminates, foams, films and films under specified conditions.
Standards:
ASTM D 2863, ISO 4589
Main features:                       
1, The oxygen concentration measurement range: 0~100%;
2, The oxygen concentration in accuracy of ± 0.1%
3, N2 mass flow controller range:1.0~15.0L/min;Accuracy:2%
4, O2 mass flow controller range:1.0~15.0L/min;Accuracy:2%
5. Oxygen sensor response time: <10s;
6.The gas flow rate in combustion tube is 40mm / s ± 2mm /s;
7, The gas flow rate in top of the combustion tube is 90mm ± 10mm/s;
8.The portable ignition flame can be adjusted.
9.The measurement time is 5mins, and the accuracy is ± 0.5s.10.Pressure meter accuracy is ± 1%;
11, Burning tube inner diameter:   80mm
12.Burning tube height: 500mm
13.Working pressure:0.1Mpa;
Main Features:
1.Equipped with a high temperature quartz glass tube, the outlet inner diameter is 40mm;
2.The rotor flowmeter regulates the flow of oxygen and nitrogen gas.
3.The gas flow mixing control unit is equipped with N2 and O2 mass flow controller to control the gas flow accurately.
4.Equipped with special ignition; portable and easy to operate
Control System:
1.The system automatically adjusts the oxygen concentration after the value is set;
2.The key electrical components are imported which are accurate, reliable and durable;
3.With a safety protection circuit;

 
ModelFTech-ISO 4589
Dimension550(L)×435(W)×670(H)mm
Packing dimension710(W)×500(L510(H)mm
PowerAC 220V, 50/60Hz, 5A
Weight30kg