Anesthesia Machine with Ventilator Scope Adults Children

China Anesthesia Machine with Ventilator Scope Adults Children, Find details about China Adults Children Ventilation Mode, Ventilator Children from Anesthesia Machine with Ventilator Scope Adults Children

Model NO.
AJ-2102
Anesthesia
Semi Open Suction
Type of Operating Light
Single Operating Light
Low Oxygen Pressure Alarm
Oxygen Pressure < 0.2MPa
Flow Meter
0.2:0.1~10L/Min N2o:0.1~10L/Min
Rapid Oxygen Flow
35~75L/Min
Ventilation Mode
Ippv Sippv Plv Peep Manu
Trademark
ANJUE
Transport Package
Wooden Case
Specification
L940*W790*H1270MM
Origin
China, Anhui
HS Code
9018907010
Model NO.
AJ-2102
Anesthesia
Semi Open Suction
Type of Operating Light
Single Operating Light
Low Oxygen Pressure Alarm
Oxygen Pressure < 0.2MPa
Flow Meter
0.2:0.1~10L/Min N2o:0.1~10L/Min
Rapid Oxygen Flow
35~75L/Min
Ventilation Mode
Ippv Sippv Plv Peep Manu
Trademark
ANJUE
Transport Package
Wooden Case
Specification
L940*W790*H1270MM
Origin
China, Anhui
HS Code
9018907010
AJ-2102 Anesthesia Machine, designed with state-of-the-art technologies, is one of high utility with excellent performance and complete functions. Both outstanding and inexpensive, it has the best cost performance that can satisfy any customer.
  1. The Ventilator adopts microcomputer control with high-definition 5.7" LCD display, integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.
  2. The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose one kind of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
  3. Reliable pneumatic transmission system.
  4. Economic and proper configuration of technologies, optional as required by the Customer.
  5. Modularized design and upgrading. Upgrading is available as required by the Customer.

TECHNICAL PARAMETERS
Part of Anesthesia Machine
Main body                 High-strength engineering plastic rack, light,
beautiful and corrosion resistant
Scope of application          adult, child
Gas source                  O2: 0.27~0.55MPa  N2O: 0.27~0.55MPa
Flow meter                   O2: 0.1~10L/min    N2O: 0.1~10L/min
O2, N2O linkage and N2O Stopper
 When using nitrous oxide, oxygen concentration>25%;
 When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off.
Flow rate of rapid oxygen supply  25~75L/min
Low oxygen pressure alarm     There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer                   it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
Respiratory circuit             working mode: all-close, semi-close, semi-open
                             releasing pressure: 0.5~7 kPa
Respiratory bellows           bellows for adults, bellows for children
tidal volume range:0~1,500 ml

Part of Ventilator
Display mode                High-definition 5.7" LCD screen display
Ventilation mode             IPPV,SIPPV,SIMV,PEE,MANUAL,SIGH
Ventilation parameters         Tidal volume          50~1500ml
Rate                 2~99bmp
SIMV rate                 2~20bpm
I:E                  2:1~1:8
Inspiratory trigger pressure -1.0~2.0 kPa 
PEEP                0~2.0 kPa
Pressure range         1.0~6.0 kPa
SIGH    1.5 times the inspiratory time
Parameters for ventilation monitoring  tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure
Monitoring of oxygen concentration  (21%~100%)
Security Alarm System
Oxygen Concentration alarm    upper limit setting range (50%~100%)
low limit setting range  (15%~50%)
Airway pressure alarm         upper limit setting range  1.0~6.0 kPa
low limit setting range   0.4~2.0 kPa
Per-minute ventilation volume alarm  upper limit setting range  3.0~30L/min
low limit setting range   1.0~10L/min
Sustained high-pressure alarm     it will give alarm when stress have consistently
been higher than 2.5 kPa 
Suffocation alarm             it will give sound and light alarm if there is
no tidal volume input for 15 seconds
Power alarm
  • Power                       AC 220V±10%     50Hz±1Hz
                            (UPS, Storage battery)
*Items within brackets are optional.