China Fully Automatic Sample Digestion System Rotary Oscillating and Mixing Technology, Find details about China Sample Digestion System, Automatic Digestion System from Fully Automatic Sample Digestion System Rotary Oscillating and Mixing Technology
Rotary oscillating and mixing technology - rotary oscillating and mixing has large amplitude and good mixing effect, and the swing amplitude is 5-10 times that of reciprocating technology.
Sample non-contact mixing - avoid cross contamination and loss of samples. The instrument can be used for oscillation, mixing, and setting of speed and time during process of adding liquid, cooling, constant volume, etc.
Positive pressure with intelligent ventilation, anti-corrosion and anti-condensation water design: adopt the integral anti-corrosion material design, to ensure the long-term operation of equipment reliable.
Digestion equipment is completely enclosed: the powerful fan inhales fresh air and injects it into the control and implementation system of digestion equipment, where a positive pressure state is formed to prevent acidic gas from entering the fume hood. Acid mist of samples is wrapped by the positive pressure before going to the ventilation system.
Solve the problem of water vapor formation in the inside wall of fume hood:the reason for the formation of water vapor is the condensation phenomenon formed by the high temperature environment encountering low temperature conditions, and the key to solve this problem is to eliminate the internal and external temperature difference and remove the gas with a large water content, so that it cannot form condensation.
Intelligent internal and external environmental temperature determination, independent power supply system: solve the problem of time limit for stopping positive pressure and negative pressure ventilation.
The digestion equipment has a set of independent environmental temperature determination system and ventilation control system.
Even if the operator turns off its power switch, the system will maintain the ventilation state if there is large temperature difference between the inside and outside.
Digestion and heating system adopts partition design:
The two digestion units can be controlled and operated independently or simultaneously, and the digestion module adopts graphite heater with good temperature uniformity and high
temperature resistance.
Two independent temperature zones can be controlled and operated independently or simultaneously
Liquid supply system
Sample head-space system: the instrument is equipped with one air channel, which can be used for all the head-space
of the pre-reagent or the head-space at any position to prevent the reaction of different reagents.
The instrument is equipped with 8 reagent channels, with multiple reagent automatic switching functions, and each
reagent channel is independently controlled so that they will not affect each other.
The distilled water is supplied by independent pipeline (single module and double channel liquid adding pipeline) to
prevent reagent adding error caused by the same reagent pipeline.
Adding reagent adopts peristaltic pump with high precision and corrosion resistant, which can be calibrated at any time with high accuracy.
Instrument constant volume system
Use "high-precision ultrasonic liquid level sensor" to implement constant volume for any size digestion tube.
The finishing digestion tube cooperates with the above sensor to make the volumetric accuracy of 10ml-50ml ≤±1%.
Instrument temperature control system
Temperature control range:RT~300ºC
Temperature control accuracy:≤1ºC
Adopt PID intelligent control
Measuring range | 0.1mg ~ 240mg N |
Analysis time | 5 ~ 10min/sample |
Reproducibility | Average value relative error ≤±0.5% |
Recovery | >=99.5% |
Burette volume | 2.0μL/step |
Sample capacity | solid<=5g/sample, liquid<=20mL/sample |
Water consumption in the distillation process | 1.5L/min |
Data storage capacity | 1000 groups |
Power supply | 220VAC±10%, 50Hz |
Power | 2Kw |
Net weight | 38Kg |
Dimensions | 455mm×391mm×730mm |
Q: How about quality of the goods?
A:lAll the running products comply with CE standards.
lIndependent quality inspection team, conducting multiple inspections and random inspections from parts supply to product dispatch.
lWell-known third-party testing service available, such as SGS, BV, and TUV, etc.
Q: What are your pre- and after-sales services?
A:lTimely reply to pre-sales enquiries
lReal-time update of manufacturing and shipment status
lContracted third-party maintenance company in over 60 countries
lAssured availability of parts supply for 10 years at least
Q: Is OEM service available.
A: Most of the running products are available for OEM customization, including logo, package and function, if attainable, customization.
Q: How long is the lead time:
A: Generally, the lead time for most stocked products is about 7~15 days, vary by models and production schedule. For those non-stocked products, the manufacturing lead time is around 20 days.