China CE Approval Manufacture Medical Device Professional Mobile HD Screen Multiparameter Patient Monitor, Find details about China Patient Monitor, Patient Monitoring Equipment from CE Approval Manufacture Medical Device Professional Mobile HD Screen Multiparameter Patient Monitor
Application Range | Adult/Pediatric/Neonatal | Dimension and Weight | 175mm*140mm*300mm; 3.3kg |
Display | 7" real color TFT screen | Package | 360m*250m*300m; 4.5kg |
Operating environment | Displayed Parameters | ||
Temperature | 0~40ºC | Time | Battery-backed quartz crystal clock |
Humidity | 15%~80% | ECG | ECG Waveform Scale, displayed lead |
Pressure | 86.0kPa~106.0kPa | Heart Rate | Derived from ECG or SpO2 |
Power Requirements | NIBP | Pressure (systolic, mean and diastolic) | |
AC | 100~240V, 50Hz/60Hz | Pulse Oximeter | Pulse Rate, SpO2 waveform, and Oxygen saturation |
DC | Built-in rechargeable battery | Respiration Rate | Respiration rate derived from ECG |
Battery | 11.1V 24wh Li-ion battery | ||
2h operating after full charge | Trends | HR, RR, NIBP, TEMP and SpO2 | |
5min operating after low battery alarm | Temperatures | One channel |
ECG | SpO2 | ||
5 Leads | RA, LA, LL, RL, V | Measurement range | 0~100% |
Lead mode | I, II, III, AVR, AVL, AVF, V | Resolution | 1% |
Gain | x1.25mm/mV, 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV (error≤5%) | Accuracy | 70%~100%±2% 40%~69%±3% |
Sweep speed | 6.25mm/s, 12.5mm/s, 25mm/s , 50mm/s(error≤±10%) | Pulse Rate | |
Bandwidth | Surgery 1~25Hz | Measure/alarm range | 20~300bpm |
Monitor | 0.5~40Hz | Resolution | 1bpm |
Diagnostic | 0.05~130Hz | Accuracy | ±3bpm |
Sensitivity | >200μpVP-P | RESP | Oscillometry |
CMRR | ≥112dB | Method | Impedance between RA-ALL |
Differential Input Impedance | >5M | RESP impedance range | 0.3~3Ω |
ohm Electrode Offset Potential | ±300mV | Base Impedance range | 200Ω~4000Ω |
Polarization Voltage | ≥±500mV | Bandwidth | 0.3~2.5Hz |
Input Impedance | ≥5M | Gain | ×0.25, x 0.5, x 1, x 2, x 4 |
Calibration Voltage | 1mV±5% | RESP rate | |
ST Segment | Adult | 0~120BrPM | |
ST Segment Range | -2.0mV~+2.0mV | Neonatal/Pediatric | 0~150BrPM |
Accuracy | 0.02mV | Accuracy | 7~150BrPM(±2BrPM) 0~6BrPM(unspecified) |
NIBP | Apnea Alarm | 10~40s | |
Method | Oscillometry | TEMP (optional) | |
Measure mode | Manual, Auto, STAT | Channel | 1 |
Measure interval in auto mode | 1~5/10/15/30/60/90/120/180/240/480min | Measure &alarm range | 0~50ºC |
Measure period in STAT mode | 5min | Resolution | 0.1ºC |
Pulse rate range | 40~240bpm | Accuracy(no sensor) | ±0.1ºC(0~50ºC) |
Measure & alarm range | Standard accessories | NIBP cuff & tube | |
Adult: | SYS 40~280mmHg | ||
DIA 10~220mmHg | ECG cable & electrodes | ||
MEAN 20~240mmHg | SpO2 sensor | ||
Pediatric | SYS 40~220mmHg | Lithium-ion battery | |
DIA 10~160mmHg | Power cable | ||
MEAN 20~170mmHg | Operator's manual | ||
Neonatal | SYS 40~135mmHg | ||
DIA 10~100mmHg | Optional accessories | TEMP probe (skin) | |
MEAN 20~110mmHg | |||
Static pressure measurement | Trolley bracket | ||
Range | 0~300mmHg | Hanging Bracket | |
Accuracy | ±3mmHg | Monitor recorder | |
Resolution | 1mmHg Accuracy | Vehicle-mounted power supply |
HL | About 3-5 working days Good at Asia,America,Europe,etc |
FedEx | About 3-7 working days Good at India,America,Europe,etc |
UPS | About 3-6 working days Good at America,Europe,South Africa,etc |
TNT | About 3-8 working days Good at The middle east,Africa |
EMS | About 15-20days Good at The Russian Federation |
SF express | About 5-10 working days Good at America,Asia,etc |
By Air | About 4-7 working days Most country available |
By Sea | About 7-30 days Only available for coastal areas.Cheapest and longest |