China Hospital Equipment Bedside Etco2 Patient Monitor, Find details about China Bedside Patient Monitor, Multi-Parameter Patient Monitor from Hospital Equipment Bedside Etco2 Patient Monitor
Description | Multiparameter Etco2 Patient Monitor |
Model | CMS8000 |
Brand | CONTEC |
Certificate | CE/FDA |
ECG | Lead mode: 3-lead or 5-lead Lead selection: I, II, III, aVR, aVL, aVF, V Waveform: 5-lead, dual-channel 3-lead, single-channel Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s |
HR: | Measurement and alarm range: 15~350 bpm Accuracy: ±1 % or ±1 bpm, whichever is greater Alarm accuracy: ± 2 bpm Resolution: 1 bpm ST-segment monitoring: Measurement and alarm range: -2.0 mV ~ +2.0 mV Accuracy: -0.8 mV~+0.8 mV ±0.04 mV or ±10%, whichever is greater Other range: unspecified Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC Pacemaker: yes |
RESP | Method: R-F(RA-LL) Impedance Respiration rate: Measurement and alarm range: 0~150 rpm Resolution: 1 rpm Measurement accuracy: ±2 rpm Alarm accuracy: ±3 rpm Apnea alarm: 10~40s Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s |
NIBP | Method: Oscillometry Mode: Manual/AUTO/STAT Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes Measurement period in STAT mode: 5 minutes Measurement and alarm range: 10 ~ 270 mmHg Resolution: 1 mmHg Cuff pressure accuracy: ±3 mmHg Measurement accuracy: Maximal mean deviation: ±5 mmHg Maximal standard deviation: 8 mmHg Over-pressure protection: -Adult mode: 297±3 mmHg -Pediatric mode: 240±3 mmHg -Neonatal mode: 147±3 mmHg |
SpO2 | Measurement and alarm range: 0 ~ 100% Resolution: 1% Measurement accuracy: 70%~100%: ±2%; 0%~69%: unspecified PR Measurement and alarm range: 30 ~ 250 bpm Measurement accuracy: ±2 bpm or ±2%, whichever is greater |
TEMP | Channel: dual-channel Measurement and alarm range: 0 ~ 50ºC Resolution: 0.1 ºC Accuracy: ±0.1 ºC |
EtCO2 | Method: Sidestream or Mainstream Measuring Range: 0~150mmHg Resolution: 0~69 mmHg, 0.1 mmHg 70~150 mmHg, 0.25 mmHg Accuracy: 0~40 mm Hg ±2 mm Hg 41~70 mm Hg ±5% 71~100 mm Hg ±8% 101~150 mm Hg ±10% AwRR Range: 2~150 rpm AwRR Accuracy: ±1BPM Apnea Alarm: Yes |
IBP | Channel: dual-channel Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2 Measuring and Alarm Range: -10~300 mmHg Resolution 1 mmHg Accuracy: ±2% or 1mmHg, whichever is greater Power supply: 100-240V~, 50/60Hz Safety classification: Class I, type CF defibrillation-proof applied part |