China Ce Approved ICU Room Advanced Type Anesthesia Machine Ljm 9900, Find details about China Anaesthesia Machine, Anesthesia Machine from Ce Approved ICU Room Advanced Type Anesthesia Machine Ljm 9900
Physical&Environment Specification | |
Dimension(WxHxD) | 970*920*1670 mm |
Weight | 95kg |
Working temperature | +5ºC~+40ºC |
Working humidity | ≤80% |
Storage temerature | -20ºC~+50ºC |
Storage humidity | ≤93% |
Power supply | |
Main power | 100-240V 50/60Hz |
Battery | not less 8 hours |
Outlets | 4 on back, 1.5A individual |
Pneunatic Specification | |
Gas supply | Air, O2, N2O |
Flowmeter: O2 range | 0-1L/min, 1-10 L/min |
N2O range | 0-1L/min, 1-10 L/min |
Air range | 0-1L/min, 1-10 L/min |
Auxiliary O2 range | 0-1L/min, 1-10 L/min |
Gas system:Oxygen supply failure alarm | Yes |
N2O-cut-off system | Yes |
O2 flush | 25-75L/min |
ACGO | optional |
AGSS | optional |
Breathing system specification | |
Operation model | Closed |
CO2 Absorbed container | 1.6L |
APL range | Adjustable |
Material | ABS |
CO2 bypass | Optional |
Ventilator Specification | |
Driven mode | pneumatically driven and electronically controlled |
Patient type | Adult Pediatric infant |
Ventilator mode | IPPV(VC),IPPV(PC),SIPPV(VC),SIPPV(PC).IMV(PC),IMV(VC),SIMV(PC),SIMV(PC),SIGH,PEEP,MANUAL STANDBY |
Setting mode | Button & encoder |
Tidal volume | 10-1500ml |
Ventilation frequency | 5-100 bpm |
I:E ratio | 2:1-1:8 |
Pressure(support) range(Only in PSV) | NO |
Pressure(inspired) range | 1-60 cmH2O |
PEEP | OFF, 0.1-1.5Kpa |
Tslope | NO |
Inspiratory plateau | OFF, 0-60% Ti |
Trigger sensitivity (Pressure) | -1 ~ -9cmH2O |
Compensation | Compliance and leakage compensation, fresh gas compensation, altitude compensation |
Ventilator Monitoring | |
Display | 10.4" TFT |
Pressure | Ppeak, Pplat, Pmean, PEEP |
Gas | O2 concentration, optional CO2 |
Graph display | P-T, F-T, F-V, P-V loops, optional CO2-T |
Alarm | High tidal volume/Low tidal volume |
High minute volume Low minute volume | |
Sustained high airway pressure | |
High FiO2%/Low FiO2% | |
Low gas supply | |
Low battery | |
Oxygen battery off | |
Power faliure | |
Alarm slience:110S | |
Vaporizer | Halothane,Enflurane, Isoflurane,Sevoflurane |