China Comen Cm100 One-Piece 10-Lead EKG Cable, Find details about China EKG Cable, Cardioline EKG Cable from Comen Cm100 One-Piece 10-Lead EKG Cable
Manufacturer | Model |
---|---|
Carewell | tutti i modelli / all models stand-alone |
Biolight | BLT-1203, BLT E12A, BLT E12, E70, E80 (solo/only conn. Dsub15) |
Comen | CM100, 300, 600, 1200, 1200A, 1200B |
Contec | Contec All |
Trismed | Cardipia, tutti i modelli / all models |
ECG interpretation: | Nihon Kohden ECG Analysis Program for Adults and Pediatrics |
Computerized measurements: | 12-lead analysis |
ECG analysis frequency: | 500 samples/second (sps) |
Digital sampling rate: | 2,000 samples/second/channel |
Dynamic range: | AC Differential ± 5mV, DC offset ±300 mV |
Resolution: | 4.88 μV/LSB @ 500 sps |
Frequency response: | -3 dB @ 0.01 to 150 Hz |
Low cut-off frequency: | 0.01 Hz, 0.02 Hz, 0.16 Hz or 0.32 Hz (-3 dB limits) |
High cut-off frequency: | Configurable at 20 Hz, 40 Hz, 100 Hz or 150 Hz |
Common mode rejection: | >100 dB (with AC filter switched on) |
Input impedance: | >10MΩ @ 10 Hz, defibrillator protected |
Patient leakage: | <10 μA |
Heart rate meter: | 30 to 300 BPM ±10% or ±5 BPM, whichever is greater. |
Start-up time: | Less than 7 seconds |
Temperature Operating: | 5°C to 40°C | Transport/storage: | -15°C to 50°C |
Humidity Operating: | 25% to 95% RH non-condensing; | Transport/storage: | 25% to 95% RH non-condensing |
Pressure Operating: | 700 to 1060 hPA | Transport/storage: | 500 to 1060 hPA |
P/N | Description | Connector Imagine |
MPK014A | IEC Din3.0 No resistance | |
MPK014B | AHA Din3.0 No resistance | |
MPK014C | IEC Banana No resistance | |
MPK014D | AHA Banana No resistance | |
MPK014E | IEC Snap No resistance | |
MPK014F | AHA Snap No resistance | |
MPK014G | IEC Clip No resistance | |
MPK014H | AHA Clip No resistance |