Automatic Closed Cup Flash Point Tester

China Automatic Closed Cup Flash Point Tester, Find details about China Close Cup Flash Point Test, ASTM D93 from Automatic Closed Cup Flash Point Tester

Model NO.
Hk-261B
Cooling Way
Compelling Cooling by Double Fan
Testing Method
High Precision Flame Ionization Test
Ambient Requirement
Temperature 15~30°c, Humidity: ≤85%
Trademark
HK
Transport Package
Wooden Case
Specification
550× 310× 400(mm)
Origin
China
HS Code
90278099
Model NO.
Hk-261B
Cooling Way
Compelling Cooling by Double Fan
Testing Method
High Precision Flame Ionization Test
Ambient Requirement
Temperature 15~30°c, Humidity: ≤85%
Trademark
HK
Transport Package
Wooden Case
Specification
550× 310× 400(mm)
Origin
China
HS Code
90278099

HK-261B Automatic Pensky-Martens Closed Cup Flash Point Tester

Overview

  HK-261B Automatic Pensky-Martens Closed Cup Flash Point Tester conforms to ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to test for those liquids which have a kinematic viscosity of 5.8mm 2/s (cSt) or more at 37.8ºC or 9.5mm 2/s (cSt) or more at 25ºC (77ºF), or that contain suspended solids, or have a tendency to form a surface film while under test. Other classification flash points have been estabilished by these departments for liquids using these test methods.

 

Features

1.Touch screen control, it can connect to the computer

2.Indenpendent heating module with monitoring interface, it follows the standard requirement to heat 3.Automatic ignition,flash point detecting alarm, calculate, print, record and air cooling

3.Standard reagent calibration:According to the standard, there are procedure A and procedure B

 

Technical Parameters

1.Rated voltage: AC220V±10% 50Hz

2.Power: 700WScreen: 4 wires resistor touch screen

3.Ignition: Gas ignition

4.Printing method: Thermo printer

5.Repetitive error: ±1ºC

6.Ambient requirement: Temperature: 10~40ºC; Humidity≤85%