GD-261A ASTMD93 Closed Cup Flash Point Tester (Pensky-Martens Closed Cup)
Product Description:
» GD-261A ASTMD93 Closed Cup Flash Point Tester is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester.
» GD-261A ASTMD93 Closed Cup Flash Point Tester is used to make determination of the closed cup flash point of the petroleum products.
Main Features:
1. Simulation tracking displays function curves of heating up and test time.
2. Equipped with the English prompt function of mis-operation.
3. Equipped with prompt function of test date, test time and other parameters.
4. Automatically calibrate the effect of atmospheric pressure to test and calculate the corrected value.
5. Differential coefficient detection. Automatically correct the system deviation.
6. Automatically open the lid, detect the flash point and print test data.
7. The test arm rises up and lowers down automatically.
8. Igniting with electric igniting gun, forced air cooling.
Main Specifications:
1 | Power supply | AC(220±10%)V, 50Hz |
2 | Temperature measurement | Range: Room temperature~300°C |
Repeatability: ≤ 2°C |
Reproducibility: ≤ 4°C |
Accuracy: 0.1°C |
3 | Heating rate | Procedure A: (5~6)°C/min |
Procedure B:(1~1.6)°C/min |
Automatic control and manually adjustable. |
4 | Stirring rate | (1) Procedure A: (90~120)RPM |
(2) Procedure B: (250±10)RPM |
Automatic control and manually adjustable. |
5 | Igniting mode | (1) Igniting source: gas(or other civilian fuels,the same below) |
(2) Electric ignition. Gas flame diameter:3.2mm~4.8mm |
6 | Ambient temperature | ≤ 40°C |
7 | Relative humidity | ≤ 80% |
8 | Maximum power consumption | 350W |