Fast Auto Electron Precision Electronic Single Channel Pipette Automatic Electronic Micro Pipette

China Fast Auto Electron Precision Electronic Single Channel Pipette Automatic Electronic Micro Pipette, Find details about China Pipettes, Pipette from Fast Auto Electron Precision Electronic Single Channel Pipette Automatic Electronic Micro Pipette

Model NO.
HT-0316
Volume Range(UL)
0.1-2.5 and Others
Trademark
Dragon
Transport Package
Color Box with Export Carton
Specification
CE, ISO13485
Origin
Beijing, China
HS Code
9018909000
Model NO.
HT-0316
Volume Range(UL)
0.1-2.5 and Others
Trademark
Dragon
Transport Package
Color Box with Export Carton
Specification
CE, ISO13485
Origin
Beijing, China
HS Code
9018909000
Mechanical Pipettes (Adjustable & Fixed)
All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22ºC.

 
  


 
Features for MicroPette
- Lightweight, ergonomic, low force design
- Digital display clearly reads volume setting
- The pipettes cover volume range of 0.1μl to 10ml
- Easy to calibrate and maintain with tool supplied
- Design helps avoid repetitive strain injuries
- Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
- The low part is available for autoclaving
- Manufactured from innovative materials

Features for multi-channel MicroPette
- 8 and 12 channel pipettes are available for standard 96-well plate
- Dispensing head rotates for optimum pipetting convenience
- Individual piston and tip cone assemblies allow easy repair and maintenance
- Compound material tip cone design allows visual seal verification
- Can be used with universal style pipette tips
  
Calibration

Please visit our website for online calibration at hiprove.en.made-in-china.com. Through the online calibration software, users can perform simple, accurate and timely calibration with free cost, and avoid the calibration error due to non-professional operation.
 
We will support and help you to achieve consistently excellent results.
 





Specifications

 
  
Single-channel Adjustable Volume Pipettes
Volume RangeIncrementTest VolumeMaximum permissible systematic error (Inaccuracy)Maximum permissible random error (Imprecision)
%μL%μL
0.1-2.5μL0.05μL2.5μL2.50%0.06252.00%0.05
1.25μL3.00%0.03753.00%0.0375
0.25μL12.00%0.036.00%0.015
0.5-10μL0.1μL10μL1.00%0.10.80%0.08
5μL1.50%0.0751.50%0.075
1μL2.50%0.0251.50%0.015
2-20μL0.5μL20μL0.90%0.180.40%0.08
10μL1.20%0.121.00%0.1
2μL3.00%0.062.00%0.04
5-50μL0.5μL50μL0.60%0.30.30%0.15
25μL0.90%0.2250.60%0.15
5μL2.00%0.12.00%0.1
10-100μL1μL100μL0.80%0.80.15%0.15
50μL1.00%0.50.40%0.2
10μL3.00%0.31.50%0.15
20-200μL1μL200μL0.60%1.20.15%0.3
100μL0.80%0.80.30%0.3
20μL3.00%0.61.00%0.2
50-200μL1μL200μL0.60%1.20.15%0.3
100μL0.80%0.80.30%0.3
50μL1.00%0.50.40%0.2
100-1000μL5μL1000μL0.60%60.20%2
500μL0.70%3.50.25%1.25
100μL2.00%20.70%0.7
200-1000μL5μL1000μL0.60%60.20%2
500μL0.70%3.50.25%1.25
200μL0.90%1.80.30%0.6
1000-5000μL50μL5000μL0.50%250.15%7.5
2500μL0.60%150.30%7.5
1000μL0.70%70.30%3
2-10mL0.1mL10mL0.60%600.20%20
5mL1.20%600.30%15
2mL3.00%600.60%12
8-channel Adjustable Volume Pipettes
Volume RangeIncrementTest VolumeMaximum permissible systematic error (Inaccuracy)Maximum permissible random error (Imprecision)
%μL%μL
0.5-10μL0.1μL10μL1.50%0.151.50%0.15
5μL2.50%0.1252.50%0.125
1μL4.00%0.044.00%0.04
5-50μL0.5μL50μL1.00%0.50.50%0.25
25μL1.50%0.3751.00%0.25
5μL3.00%0.152.00%0.1
50-300μL5μL300μL0.70%2.10.25%0.75
150μL1.00%1.50.50%0.75
50μL1.50%0.750.80%0.4
12-channel Adjustable Volume Pipettes
Volume RangeIncrementTest VolumeMaximum permissible systematic error (Inaccuracy)Maximum permissible random error (Imprecision)
%μL%μL
0.5-10μL0.1μL10μL1.50%0.151.50%0.15
5μL2.50%0.1252.50%0.125
1μL4.00%0.044.00%0.04
5-50μL0.5μL50μL1.00%0.50.50%0.25
25μL1.50%0.3751.00%0.25
5μL3.00%0.152.00%0.1
50-300μL5μL300μL0.70%2.10.25%0.75
150μL1.00%1.50.50%0.75
50μL1.50%0.750.80%0.4
Fixed Volume Pipettes
Volume RangeIncrementTest VolumeMaximum permissible systematic error (Inaccuracy)Maximum permissible random error (Imprecision)
%μL%μL
5μL-5μL1.3%0.0651.2%0.06
10μL-10μL0.8%0.080.8%0.08
20μL-20μL0.6%0.120.5%0.1
25μL-25μL0.5%0.1250.3%0.075
50μL-50μL0.5%0.250.3%0.15
100μL-100μL0.5%0.50.3%0.3
200μL-200μL0.4%0.80.2%0.4
250μL--250μL0.4%1.00.2%0.5
500μL-500μL0.3%1.50.2%1.0
1000μL-1000μL0.3%3.00.2%2.0
2000μL-2000μL0.3%6.00.15%3.0
5000μL-5000μL0.3%150.15%7.5