China Manual Type Open Cup Flash Point Testing Equipment ASTM D92, Find details about China Flash Point Measuring Instrument, Flash Point Apparatus Manual from Manual Type Open Cup Flash Point Testing Equipment ASTM D92
Introduction:
Semi-automatic Lab Equipment Cleveland Open Cup Flash Point Tester is the determination of the flash point and fire point of petroleum products by a manual Cleveland open cup apparatus.
It adopts technology of single chip micro-computer and LCD screen. English man-machine interface. The heating device is continuously adjustable. Easy to set parameters.
Specifications:
Product name | GD-3536-1 Semi-automatic Lab Equipment Cleveland Open Cup Flash Point Tester(ASTM D92) |
Power supply | AC (220±10%) V, 50Hz |
Heating device | Electric furnace heating, no naked fire, explosion prevented.The power is adjustable from 0W to 600W. The max heating temperature can reach 400ºC |
Temperature control | Single chip micrºComputer.The heating rate can meet |
Requirements of standards | GB/T 3536-2008 and ASTM D92 |
Temperature display | LCD shows temperature parameters |
The display range and accuracy | 0ºC~ 400ºC, 0.1ºC |
Flash point detecting device | It applies the test flame automatically |
Temperature sensor | RTD, PT100 |
Igniting device | (1) Ignition source: coal gas (or civil gas) (2) Flame diameter is 3.2mm~ 4.8mm |
Ambient temperature | (-10 ~ 50)ºC |
Relative humidity | ≤85% |
Maximum power consumption | 650W |
Dimension | 340*320*450mm (Temperature sensor is included) |
Accessories:
No. | Name | Unit | Quantity | Remark |
1 | Main unit | Set | 1 | |
2 | Cleveland oil cup | Piece | 1 | We can provide more if you required. |
3 | Ignition device | Set | 1 | |
4 | Thermometer | Piece | 1 | -6°C~400 °C, graduated scale is 2 °C |
5 | Thermometer support | Set | 1 | |
6 | Power supply wire (250 V 6 A) | Piece | 1 | |
7 | Fuse 5 A (Φ5×20) | Piece | 2 | |
8 | Latex tube Φ6 | Meter | 1.5 | Could not be used as coal gas tube |
9 | Joint for coal gas tube | Piece | 1 | Connect coal gas tube and latex tube |
10 | Extinguishing cover | Piece | 1 |
GD-3536 Series ASTM D92 CLEVELAND OPEN CUP FLASH POINT TESTER
Parameter | GD-3536 | GD-3536-1 | GD-3536A | GD-3536D |
Temperature range | Ambient temperature~400ºC | Ambient temperature~400ºC | Ambient temperature~400ºC | Ambient temperature~400ºC |
Ignition mode | Ignite once manually | Ignite once manually | Ignite once manually | Automatic |
Temperature display | Thermometer | LCD | Color LCD | Color LCD |
Printing | No | No | Yes | Yes |
Communication port | No | No | No | Yes |
Other function | Manually adjust heating rate | Automatically control temperature and display by LCD | Automatic | Automatic, with RS232 communication port |
Dimension | 350*290*350mm | 340*320*450mm | 410*360*420mm | 520*380*310mm |