Veterinary Monitor (Moni 8000DV)

China Veterinary Monitor (Moni 8000DV), Find details about China Animal Monitor, Pet Monitor from Veterinary Monitor (Moni 8000DV)

Model NO.
Moni 8000DV
Trademark
SHINOVA
Specification
ISO, CE
Origin
China
HS Code
90181930
Model NO.
Moni 8000DV
Trademark
SHINOVA
Specification
ISO, CE
Origin
China
HS Code
90181930
Veterinary Monitor
(Multi-parameter Monitor for Veterinary)
Model: Moni 8000DV
 
Performance Specifications
Display:    12.1 inch color TFT (diagonal)
Rolling and refreshing waveform display
Resolution: 800x600
Multi displays selectable
Trace:        7 waveforms
Sweep Speed: 6.25mm/s,12.5mm/s, 25mm/s, 50mm/s
Indicator:  Power indicator light
Battery indicator light
Inter face: Parameter cable inter face
AC Power input socket
Network inter face
External VGA inter face
for connection of an Alternate Display
Battery:      Lead Acid /Rechargeable
Maximum 12 hours for charging;
 4 hours for continuous working
Trend time:  1~120 hours
Alarm:     User-adjustable High and Low limits audible and visual alarm
Networking: Connected to central monitoring system
Recorder:     Built-in, thermal array
Waveform:   2 channels
Record mode: manual, on alarm, time -defined, etc Paper width: 50mm
Print speed: 12.5 mm/s, 25mm/s, 50mm/s
 
 
ECG
5-lead and 3-lead selectable
Input:      5 leadwire cable: RA; LA; RL; LL; V or R; L; N; F; C
3 leadwire cable: RA; LA; LL or R; L; F
Lead selection:  5-lead: I; II; III; aVR; aVL
3-lead: I; II; III
ECG waveform: 2 channels
Gain selection:  x0.25; x0.5; x1; x2; auto
Sweep speed:   12.5mm/s, 25mm/s, 50mm/s
Heart Rate range: 15-350bpm;
Accuracy:      ±1bpm or ±1%, whichever is greater
Resolution:    1bpm
Protection:     Withstand 4000VAC/50Hz voltage in isolation
Against electrosurgical interference and defibrillation
CMRR:    Diagnostic Mode :> 90dB
Monitoring Mode :> 105Db
 Operation Mode :> 105dB
Calibration signal:   1mv±5%
Operation mode:     Diagnose, monitoring operation
Alarm Range:       15-350bpm
S-T protection:      Yes
Measurement range: -2.0mV~2.0mV
Alarm Range: -2.0mV~2.0mV
Arrhythmia analysis: Yes
Alarm:       Yes audible and visual alarm, alarm events recallable
 
Respiration
Method:  RA-LL impedance
Measurement range: 6-150 rpm;
Resolution:    1rpm
Apnea alarm:  YES
Alarm:            YES, audible and visual alarm, alarm events recallable
 
NIBP
Method:  Automatic oscillometric
Operation modes: Manual/Automatic/Continuous
Measurement unit:    mmHg/kPa selectable
Measurement types: Systolic, Diastolic, Mean
Auto measure time:   Adjustable
Measurement range: 10-270mmHg
 
Accuracy of blood pressure measurement:   The mean error shall be less than ±5 mmHg
The standard deviation shall be less than 8 mmHg
Over-pressure protection: Double safety protection
Resolution:    1mmHg
Alarm:            Systolic, Diastolic, Mean
 
Nellcor SpO2 VetSat(Optional)
Measurement range: 0~100%
Resolution:  1%
Accuracy:      ±2% (70-100%, MAX-A, MAX-AL, MAX-N, MAX-P,
MAX-I and MAX-FAST sensors);
±3% (70-100%, D-YS, DS-100A,
OXI-A/N and OXI-P/I sensors);
 0-69% unspecified
Alarm range:  0~100%
Pulse rate:     Range: 20~250bpm
Resolution: 1bpm
Accuracy: ±3bpm
Unspecified   Alarm range: 20~250bpm
 
Temperature
Accuracy:      ±0.1ºC(excluding the sensor)
Measurement range:   0~50ºC
Channel: 1-channel. Provide
Resolution:    0.1ºC
 
IBP (Optional)
Measurement Mode:     Pressure Transducer
Measurement Types:    Systolic, Diastolic, Mean
Channel: 1 channel
Pressure names:   AVP, CVP, PAR, LAP, RAP, ICP
Measurement Range:    -10mmHg~300mmHg
IBP Accuracy:              0~100mmHg  ±4mmHg
100mmHg~300mmHg  ±2%
Pulse Accuracy:               ±1% or ±1bpm, whichever is greater
Calibration Mode:        Zero Calibration
 
Standard Configurations: ECG, RESP, NIBP, SpO2, 1-TEMP
Optional Configurations:  1-IBP, Thermal Recorder
 
EtCO2 (Optional)
 
Mainstream CO2(Optional)
Method: Infrared Absorption
Measuring mode: Mainstream
Measurement range: 0 -150 mm Hg, 0 to 79%, 0 to 20kPa (at 760mmHg)
Resolution:    0.1 mmHg 0 to 69 mmHg
0.25mmHg 70 to 150mmHg
Accuracy:      ± 2 mm Hg (0-40 mm Hg)
± 5% of reading (41-70 mm Hg)
± 8% of reading (71-100 mm Hg)
± 10% of reading (101-150 mm Hg)
Alarm range: Same as Measurement rang
 
 
Safety
CE marking according to MDD93/42/EEC
 
Dimension: 320mm×170mm×280mm
Net Weight: 6.5kg
 
Operation Environment
Power: AC100-250V, 50/60Hz.
Temperature: 0-40ºC
Humidity: 15-85