HK-0325 Greases Oxidation Stability Apparatus

China HK-0325 Greases Oxidation Stability Apparatus, Find details about China Oxidation Stability Grease, ASTM D942 Tester from HK-0325 Greases Oxidation Stability Apparatus

Model NO.
HK-0325
Test Station
2 Holes
Ambient Requirements
Temperature10~40ºC ;Humidity≤85%
Trademark
HK
Transport Package
Carton + Pallet
Origin
China
Model NO.
HK-0325
Test Station
2 Holes
Ambient Requirements
Temperature10~40ºC ;Humidity≤85%
Trademark
HK
Transport Package
Carton + Pallet
Origin
China
This apparatus measures the net change in pressure resulting from consumption of oxygen by oxidation and gain in pressure due to formation of volatile oxidation by-products. It predicts neither the stability of greases under dynamic service conditions, nor the stability of greases stored in containers for long periods, nor the stability of films of greases on bearings and motor-parts. It should not be used to estimate the relative oxidation resistance of different grease types.

Overview
HK-0325 Greases Oxidation Stability apparatus conforms to the ASTM D942 Standard Test Method for Oxidation Stability of Lubricating Greases by the Oxygen Pressure Vessel Method, This apparatus determines resistance of lubricating greases to oxidation when stored statically in an oxygen atmosphere in a sealed system at an elevated temperature under conditions of test
Features
1. All procedures controlled by a touched LCD screen, it can record and display test value, also determine final point automatically. 
2. We equip with 0.1 level pressure transmitters for the apparatus, with pressure calibration device.
3. We equip Agilent measured transmitter processor for the apparatus.
4. Accurate oxygenating connector with convenient usage.
5. The apparatus is equipped with standard heatproof vessel and vessel stand.
 
 Technical parameters
1. rated voltage:AC220V±10%  50Hz
2. heating power:1500W
3. temperature control point:99ºC
4. temperature control accuracy:±0.1ºC
5. operation method: 4 wires touched LCD 
6. sensor:Pt100(RTD)
7. Test station:2 holes
8. ambient requirements:temperature10~40ºC ;humidity≤85%