Sy- 261A Automatic Closed Cup Flash Point Tester

China Sy-261A Automatic Closed Cup Flash Point Tester, Find details about China Flash Point Tester, Closed Cup Flash Point Tester from Sy-261A Automatic Closed Cup Flash Point Tester

Model NO.
SY-261A
Ambient Temperature
≤ 40°c
Relative Humidiyt
≤ 80%
Maximum Power Consumption
350W
Gas Flame Diameter
3.2mm~4.8mm
Temperature Measurement Range
Room Temperature~300°c
Trademark
SAMYON
Transport Package
Carton Box
Origin
Cn
HS Code
9031809090
Model NO.
SY-261A
Ambient Temperature
≤ 40°c
Relative Humidiyt
≤ 80%
Maximum Power Consumption
350W
Gas Flame Diameter
3.2mm~4.8mm
Temperature Measurement Range
Room Temperature~300°c
Trademark
SAMYON
Transport Package
Carton Box
Origin
Cn
HS Code
9031809090

SY-261A Automatic Pensky-Martens Closed Cup Flash Point Tester


Sy-261A Automatic Closed Cup Flash Point Tester

Sy-261A Automatic Closed Cup Flash Point Tester

Sy-261A Automatic Closed Cup Flash Point Tester

Sy-261A Automatic Closed Cup Flash Point Tester

 

Summary

This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products.

 

I. Main technical features

1. Simulation tracking displays function curves of heating up and test time.

2. Equipped with the English prompt function of mis-operation.

3. Equipped with prompt function of test date,test time and other parameters.

4. Automatically calibrate the effect of atmospheric pressure to test and calculate the corrected value.

5. Differential coefficient detection.Automatically correct the system deviation.

6. Automatically open the lid,detect the flash point and print test data.

7. The test arm rises up and lowers down automatically.

8. Igniting with electric igniting gun,forced air cooling.

 

II. Main technical specifications

1. Power supply:AC(220±10%)V,  50Hz.

2. Temperature measurement:

Range: Room temperature~300ºC.                                                      

Repeatability: ≤ 2ºC                                                      

Reproducibility: ≤ 4ºC                                                      

Accuracy: 0.1ºC

3. Heating rate:

Procedure A: (5~6)ºC/min

Procedure B:(1~1.6)ºC/min

Automatic control and manually adjustable.

4. Stirring rate:

(1) Procedure A: (90~120)RPM                           

(2) Procedure B: (250±10)RPM

Automatic control and manually adjustable.

5. Igniting mode:

(1) Igniting source: gas(or other civilian fuels,the same below)

(2) Electric ignition. Gas flame diameter:3.2mm~4.8mm

6. Ambient temperature:≤ 40ºC

7. Relative humidity:≤ 80%

8. Maximum power consumption: 350W

Main technical specifications

1. Power supply:AC(220±10%)V,  50Hz.

2. Temperature measurement:

Range: Room temperature~300ºC.                                                      

Repeatability: ≤ 2ºC                                                      

Reproducibility: ≤ 4ºC                                                      

Accuracy: 0.1ºC

3. Heating rate:

Procedure A: (5~6)ºC/min

Procedure B:(1~1.6)ºC/min

Automatic control and manually adjustable.

4. Stirring rate:

(1) Procedure A: (90~120)RPM                           

(2) Procedure B: (250±10)RPM

Automatic control and manually adjustable.

5. Igniting mode:

(1) Igniting source: gas(or other civilian fuels,the same below)

(2) Electric ignition. Gas flame diameter:3.2mm~4.8mm

6. Ambient temperature:≤ 40ºC

7. Relative humidity:≤ 80%

8. Maximum power consumption: 350W