China Venezuela Popular Medical 300 Tests Automatic Chemistry Analyzer Machine, Find details about China Medical Equipment, Lab Equipment from Venezuela Popular Medical 300 Tests Automatic Chemistry Analyzer Machine
1.Precise sample and reagent aspiration |
2.Built in bubble elimination prior to washing |
3.High pressure interior probe wash |
4.Carry-over<0.05% |
1.Light-spot Flatting Technology facilitates lower reaction volume |
2.24 hour non-stop refrigeration at 4-10ºC |
3.Minimum sample volume:2ul |
4.Reagent volume:20-25ul |
5.Minimum cuvette volume:150ul |
1.Front loaded reagent compartment for easier access |
2.Easy access for routine maintenance and troubleshooting |
3.Easy to connect pipeline and cable |
1.Real-time analytical and carousel status monitoring |
1.Levy-Jennings chart and Twin-plot chart |
2.QC Out-of-Range real-time alarm |
1.Historic date recall from reagent/calibrator/control archive |
2.Intuitive software,easy historical results |
1.User-defined customizable reflexive assays |
2.Multiple reflexive criteria may apply to a single assay |
3.A single reflexive criteria may apply up to 20 assays |
System Function | |
Fully automated,discrete,random access,STAT,urine and homogeneous immunoassays,STAT sample priority | |
Throughput | Constant 300 tests/hour,Massuring principles:Absorbance Photometry,Turbidometry |
Methodology | End-point,Fixed-time,Kinetic,Single/uual/Triple/Quadruple reagent chemistries,Monochromatic/Bichromatic |
Programming | User defined profiles and calculation chemistries |
Sample Handling | |
Sample tray | 120 positions for sample primary or secondary tubes and sample cups |
Sample volume | 2~100ul,step by 0.1ul |
Sample probe | Liquid level detection,clot detection and collision protection |
Probe cleaning | Interior and exterior automatic probe washing carry-over<0.05% |
Reagent Handling | |
Reagent tray | 80 positions in refrigerated compartment (4~10ºC) |
Reagent volume | R1:150~300ul,R2:20~150ul |
Reagent probe | Liquid level detection, collision protection and inventory check |
Probe cleaning | Interior and exterior automatic probe washing |
Reaction System | |
Reagent rotor | Rotating tray,81 cuvettes with automatic washing |
Cuvette | Optical length 5mm |
Reaction volume | 150~450ul |
Operating temperature | 37ºC |
Temperature fluctuation | ±0.1ºC |
Mixing system | 2 independent mixers |
Optical System | |
Light Source | Halogen-tungsten lamp |
Photometer | Reversed optics photometry |
Wavelength | 340nm 405nm 450nm 510nm 546nm 578nm 630nm 670nm |
Absorbance range | 0~4.0Abs(10mm conversion) |
Resolution | 0.0001Abs |
Control and Calibration | |
Calibration mode | Linear(one-point,two-point and multi-point),Logit-Log 5p,Spline,Factor |
Operation Unit | |
Operation system | Windows XP SP2 or SP3/ Windows 7 Home Basic |
Interface | RS-232 |
Working Conditions | |
Power Supply | 200~240V 50/60HZ 1000W or 100~120V 60HZ 1000W |
Temperature | 15~30ºC |
Humidity | 35~85% |
Water consumption | 18L/hour,De-ionized water |
Dimension | 912mm*721mm*570mm(W*D*H) |
HP-CHEM300H Automatic Chemistry Analyzer | Specifications | |
Analyze Unit | ||
Throughput | Constant 3000T/h (non-ISE) | |
Theory | Absorption spectrometry, transmission turbidimetry | |
Methodology | End-point \Fixed-time\Kinetic\Ion selective electrode | |
Test according set up, random access | ||
Reaction plate | Contain 120 reaction cuvettes | |
Maximum reaction time | 15 minutes | |
Minimum reaction volume | 150 ul | |
Material of cuvettes | Special UV-transparent optical glass | |
Cuvettes use | Automatically cuvettes washing | |
Optical Unit | ||
Light Source | Halogen-tungsten lamp | |
Photometer | Reversed optics | |
Wavelength | 340nm-670nm wavelength option | |
Absorption | 0-4.0 Abs(10mn conversion) | |
Reagent Unit | ||
Reagent plate | 80 reagent positions ,support for single and double reagent, reagent completely open | |
Reagent storage temperature | 4~10 degree | |
Reagent bottle | 15ml, 20ml, 30ml, 50ml or 70ml ,dead volume is less than 1.0 ml | |
Reagent volume | R1:150-300 ul,R2:20-150ul step by 1.0 ul | |
Reagent/Sample Probe | Automatically checked ;Patent technology liquid level detection;Collision protection probe cleaning:Automatic washing for both inside and outside. | |
Sample Unit | ||
Sample plate | 120 sample positions | |
Sample type | Human (animal) serum \plasma\urine etc. | |
Samples cup | Support one-off micro cups | |
Sample volume | 2-100 ul, step by 0.1 ul | |
Sample/reagent probe | Automatically checked ;Patent technology liquid level detection;Collision protection probe cleaning:Automatic washing for both inside and outside. | |
Temperature control Unit | ||
Heating method | Direct thermal solid thermostat closed | |
Temperature equilibrium time | <3 minutes | |
Reaction temperature | 37 ±0.1 degree | |
Temperature fluctuation | ±0.1 degree | |
Operation Unit | ||
Software | WindowsXP \2000 | |
Database | SQL Server | |
Data Storage | Automatic store & copy; No time and volume limit; Lifetime preservation | |
Print format | Can be set up, offer several standard format | |
Calibration mode | Linear curve type: Single-point linear, two-point linear and multi-point linear none linear; Curve type: Logistic-Log 4P,Logistic-Log5p,Spline,index automatically use best calibration curve and standard | |
Quality control methods : | Real time, within day & between day | |
Quality control diagram | L-J \cumulative\twin plot | |
Status display | Real time monitoring of reagent residual volume .Running status of reagent tray, samples tray | |
Adjustment | Dynamic and real time display of running status; Intelligent carry over settings to adjust test sequence; original reaction data record; Detail profile of alert messages | |
Setting | Operation permission setting, Item setting, cross pollution setting, cuvette blank setting | |
Others | ||
Water consumption : < 18L/h | Power Supply :AC 220V-240V 50/60Hz 1000W or 100V-200V 60HZ 1000W | |
Frequency :Maximum 600VA | Dimension :912mm×721mm×570mm | |
Net Weight : 60kg | Accessory :Software | |