Bante902-UK Benchtop Multiparameter Water Quality Meter Suitable for Measuring The pH , ORP

China Bante902-UK Benchtop Multiparameter Water Quality Meter Suitable for Measuring The pH, ORP, Find details about China Portable Water Hardness Tester, Water Quality Analyzer from Bante902-UK Benchtop Multiparameter Water Quality Meter Suitable for Measuring The pH, ORP

Model NO.
Bante902-UK
Stability Criteria
Low or High
Calibration Due Alarm
1 to 31 Days or off
Memory
Stores up to 500 Data Sets
Output
USB Communication Interface
Power
DC5V, Using AC Adapter, 220V/50Hz
Dimensions
210(L)×188(W)×60(H)mm
Trademark
Westtune
Transport Package
Packed According to Standard Goods
Specification
Weight /1.5kg
Origin
Shanghai
HS Code
9024102000
Model NO.
Bante902-UK
Stability Criteria
Low or High
Calibration Due Alarm
1 to 31 Days or off
Memory
Stores up to 500 Data Sets
Output
USB Communication Interface
Power
DC5V, Using AC Adapter, 220V/50Hz
Dimensions
210(L)×188(W)×60(H)mm
Trademark
Westtune
Transport Package
Packed According to Standard Goods
Specification
Weight /1.5kg
Origin
Shanghai
HS Code
9024102000
Bante902-UK Benchtop Multiparameter Water Quality Meter suitable for measuring the pH, ORP
    Multiparameter benchtop pH/conductivity meter contains 12 options, the meter is suitable for measuring the pH, ORP, conductivity, TDS, salinity and resistivity.




Bante902-UK Benchtop Multiparameter Water Quality Meter Suitable for Measuring The pH, ORPFunction
• Automatic temperature compensation ensures accurate readings over the entire range.
• Auto-Read function senses and locks the measurement endpoint.
• Calibration due alarm prompts user to calibrate the meter regularly.
• Setup menu allows setting the number of calibration points, stability criteria, temperature
  unit, date and time, auto-power off, etc.
• Reset function automatically resumes all settings back to the factory defaults.
• Expanded memory stores or recalls up to 500 data sets.
• USB communication interface for data transfer and timed interval readings.




The instrument includes
 Bante901/902-CN: 
•E201 pH electrode  
•CON-1 conductivity electrode 
•TP-10K temperature probe 
•pH buffer sachets 4.01/7.00/10.01 
•Conductivity standard solutions 
•84µ S/cm, 1413µ S/cm, 12.88mS/cm 
•USB cable and DC5V power adapter
Bante901/902-UK: 
•P11 glass pH electrode (Made in England)
•K10 conductivity electrode (Made in England)
•TP-10K temperature probe 
•pH buffer sachets 4.01/7.00/10.01 
•Conductivity standard solutions 
•84µ S/cm, 1413µ S/cm, 12.88mS/cm 
USB cable and DC5V power adapte


Specifications
   ModelBante902
   pH 
   Range-2.000 to 20.000pH
   Resolution0.001, 0.01, 0.1pH, selectable
   Accuracy±0.002pH
   Calibration Points1 to 5 points
   pH Buffer OptionsUSA, NIST, DIN, 2 custom buffers
   ORP 
   Range±1999.9mV
   Resolution0.1, 1mV, selectable
   Accuracy±0.2mV
   Calibration Points1 point
   Measurement ModesRelative or absolute mV
   Conductivity 
   Range0.01 to 20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm
   Resolution0.001, 0.01, 0.1, 1
   Accuracy±0.5% F.S.
   Calibration Points1 to 5 points
   Calibration Solutions10µS, 84µS, 1413µS, 12.88mS, 111.8mS
   Temperature CoefficientLinear (0.0 to 10.0%/°C), pure water
   Reference Temperature20/25°C
   Cell Constant  K=0.1, 1, 10 or custom
   TDS 
   Range0.00 to 10.00, 100.0, 1000ppm, 10.00, 200.0ppt
   Resolution0.01, 0.1, 1
   Accuracy±1% F.S.
   TDS Factor0.1 to 1.0 (default 0.5)
   Salinity 
   Range0.00 to 80.00ppt, 0.00 to 42.00psu
   Resolution0.01
   Accuracy±1% F.S.
   Resistivity 
   Range0.00 to 20.00MΩ
   Resolution0.01
   Accuracy±1% F.S.
   Temperature 
   Range0 to 105°C/32 to 221°F
   Resolution0.1°C/0.1°F
   Accuracy±0.5°C/±0.9°F
   Offset Calibration1 point
   Calibration RangeReading ±10°C
   General Specifications 
   Temperature Compensation0 to 100°C/32 to 212°F, manual or automatic
   Stability CriteriaLow or high
   Calibration Due Alarm1 to 31 days or off
   MemoryStores up to 500 data sets
   OutputUSB communication interface
   ConnectorBNC, 6-pin nimi-DIN, 3.5mm jack socket
   DisplayCustom LCD (125×100mm)
   PowerDC5V, using AC adapter, 220V/50Hz
   Dimensions210(L)×188(W)×60(H)mm
   Weight1.5kg