Superstar High Quality Nitrous Oxide System S8800 for Dental Patients

China Superstar High Quality Nitrous Oxide System S8800 for Dental Patients, Find details about China Dental Equipment, Nitrous Oxide Sedation System from Superstar High Quality Nitrous Oxide System S8800 for Dental Patients

Model NO.
S8800
Oxygen Concentration
30%-100%
N2o Concentration
0%-70%
Alarm and Protection
13 Kinds of Alarms and Protection Functions
Trademark
Superstar
Transport Package
Wooden Case
Specification
L 730*W 730*H 1445mm
Origin
Nanjing, China
HS Code
90189070
Model NO.
S8800
Oxygen Concentration
30%-100%
N2o Concentration
0%-70%
Alarm and Protection
13 Kinds of Alarms and Protection Functions
Trademark
Superstar
Transport Package
Wooden Case
Specification
L 730*W 730*H 1445mm
Origin
Nanjing, China
HS Code
90189070
SUPERSTAR High Quality Nitrous Oxide System S8800 for Dental Patients
 
10.4" TFT LCD touch screen displays the Ventilation parameters,Alarming information and Waveform.
With built-in backup power source, when outside power source goes off, back-up power source starts to work
Timing function for recording time and controlling operation
Electronic control mixed gas blender, high precision adjustment

Ventilation mode: Continuous gas supply, On-demand gas supply , Rapid oxygen supply, MANUAL
Oxygen concentration: 30%-100%
N2O concentration: 0%-70%
Alarm and protection: 13 kinds of alarms and protection functions
Oscillogram: P-T, fiow histogram

 

Gas supply

  1. Gas source                      O2(For medical use)
                                                    N2O(For medical use)
  1. Input pressure range            280 kPa~600 kPa
  2. Negative pressure gas source       flow ≥75 L/min
(when equipped with Transfer and receiving system of anaesthetic gas scavenging system)
 

Power supply

  1. Voltage                       100-240 V~
  2. Frequency                         50 HZ/60 HZ
  3. Input power                60 VA
  4. Internal battery                               DC11.1V    5200mAh
  1. Fuse                                                      
--Main power          T2AH 250V
--Internal power            T4AH 250V
 

Ventilation parameters

  1. O2 concentration       
--Adjusting range             30 %~100 %(v/v)
--Allowable error             ± 15 %, and minimum O2 concentration 30 %
  1. N2O concentration                   
--Adjusting range             0 %~70 %(v/v)
--Allowable error                        ± 15 % or ± 3 %(v/v), whichever is the
greater, and maximum  N2O concentration 70 %
  1. Continuous flow                    
--Adjusting range            0 L/min ~ 40 L/min
  1. As required flow                    
--Adjusting range            0 L/min ~ 40L/min
  1. Pressure trigger             
--Adjusting range             -1 cmH2O~ -10 cmH2O
--Allowable error              ±(1 cmH2O +10 % setting valve)
  1. Gas supply time             
--Adjusting range             1 s ~ 6 s
--Allowable error              ± 20 %
  1. Output gas flow from manual ventilation       
--Adjusting range                          0 L/min ~ 40 L/min
  1. Flow rate of oxygen flush valve         25 L/min ~ 75 L/min
  2. Maximum gross flow                          ≥ 60 L/min

Ventilation parameter Monitoring

  1. O2 concentration              
--Monitoring range                 30 % ~ 100 %
--Allowable error                    ±(2.5 % v/v+2.5 % concentration level)
  1. Airway pressure
--Monitoring range                 0 cmH2O ~ 60 cmH2O
--Allowable error                    ±( 2%full scale + 4%actual reading )
  1. Continuous flow and as required flow
--Monitoring range                 0 L/min ~ 40 L/min
--Allowable error                         >3.0 l/min:± 20 %, 3.0 l/min: ± 0.6 l/min
  1. Waveform display:
--Time- airway pressure
  1. Alarm of low voltage of back-up battery
When the voltage of back-up battery is low, the power state indication place will display the signal of "", and there will be a sound alarm.
Alarm lasting time:      >60 s
!  Warning:
If battery power fails during operation, please switch off the machine and do manual ventilation to the patient.
  1. Alarm of low pressure for input N2O
When the pressure of input N2O is under (180 ± 20) kPa, there will be a sound alarm and display of Low pressure for input N2O.
  1. Alarm of low pressure for input O2
When the pressure of input O2 is under (180 ± 20) kPa, there will be a sound alarm and display of Low pressure for input O2.
  1. High O2 alarm
When the monitoring value of O2 concentration over the upper limit of O2, there will be a sound alarm and display of O2 high.
--Setting range                            31 % ~ 100 %
  1. Low O2 alarm
When the monitoring value of O2 concentration under low limit of O2 concentration, there will be a sound alarm and display of O2 low.
--Setting range                          30 % ~ 99 %
  1. High Ppeak alarm
When the monitoring value over the upper limit of Ppeak, there will be a sound alarm and display of High Paw pressure.
--Setting range                            5 cmH2O ~ 60 cmH2O
  1. Pressure limit
When the airway pressure is over the set pressure limit, there will be a sound alarm and display of High Paw pressure, and he pressure will be limited under the set value.
--Allowable error                                                 ±5 cmH2O
  1. Protection for insufficient O2 supply    
      When O2 supply pressure is lower than 200 kPa,  NO2 shutoff valve will shut off NO2 supply.                                 
  1. Maximum limit pressure
The safety valve will be on when the Paw is over the set maximum limit pressure.
--Maximum limit pressure                                         60 cmH2O±5 cmH2O
  1. Emergency air intake valve
Emergency air intake valve will be on when the Paw is during -2cmH2O ~ -8 cmH2O. 
  1. Gas supply safety valve
Gas supply safety valve will exhaust when O2 or N2O supply pressure is over (750 ± 100) kPa.
  1. Mute
When an alarm occurs, the alarm sound may be muted by pressing AUDIO PAUSED key, but the alarm indication will display on the screen, unless the problem is solved.
--Muting time                               <120 s