Medical Device Infant Radiant Warmer for Newborn

China Medical Device Infant Radiant Warmer for Newborn, Find details about China Infant Radiant Warmer, Radiant Warmer from Medical Device Infant Radiant Warmer for Newborn

Model NO.
CNME010717
Optional Configuration
Disposable Skin Temperature Sensor, SpO2, etc
Operating Environmental Range
18°c~30°c
Baby Mode Temperature Control Range
34.5°c-37.5°c
Skin Temperature Display Range
5°c-65°c
Accuracy of Skin Temperature Sensor
±0.2°c
Temperature Uniformity of Mattress
≤2°c
Failure Alarm
Over Temperature Alarm, Deviation Alarm, etc
Trademark
CNMEDITECH
Specification
Dimension: 1270*810* 1120mm
Origin
China(Mainland)
HS Code
9018909911
Model NO.
CNME010717
Optional Configuration
Disposable Skin Temperature Sensor, SpO2, etc
Operating Environmental Range
18°c~30°c
Baby Mode Temperature Control Range
34.5°c-37.5°c
Skin Temperature Display Range
5°c-65°c
Accuracy of Skin Temperature Sensor
±0.2°c
Temperature Uniformity of Mattress
≤2°c
Failure Alarm
Over Temperature Alarm, Deviation Alarm, etc
Trademark
CNMEDITECH
Specification
Dimension: 1270*810* 1120mm
Origin
China(Mainland)
HS Code
9018909911
Medical Device Infant Radiant Warmer for Newborn 
Medical Device Infant Radiant Warmer for Newborn
Brief Introduction:
Three control Modes: Pre-warm Mode,Manual Mode and Baby Mode control by micro-computer;
Set Temperature and Baby Temperature can be displayed separately;
Horizontal angle of warmer module and inclination of the bassinet can be adjusted;
The panels surrounded bassinet can be turned outward or disassembled;
X-ray cassette under the bassinet;
APGAR timer;
Multiple failure alarm indications;
RS232 connector for data output.

Optional Configuration:
Disposable skin temperature sensor, Vertical height adjustment (VHA) stand, Weighing system and SpO2.

Standard Configuration:
CNME010717: Warmer module, bassinet, main column, mobile stand, controller, I.V. Pole and tray.
CNME010717B:
Based on CNME010717, plus the phototherapy unit.
CNME010717A:
Based on CNME010717B, plus the breath resuscitation bag; infant head fixing unit, the oxygen supplying set ( the oxygen cylinders, buoy type oxygen regulator, wall type oxygen regulator), low-pressure suction.


Environmental Requirement:
Operating range: 18°C~30°C
Ambient air velocity rate: <0.3 m/s

Medical Device Infant Radiant Warmer for Newborn Product Packaging:
Package: The whole unit is packed into one carton
Dimension: 1270mm X 810mm X 1120mm
Gross Weight: 110kg(CNME010717)
120kg(CNME010717B)
142kg(CNME010717A)


Specification:
Product name Medical Device Infant Radiant Warmer for Newborn
Power requirementAC220V-230V/50Hz or AC110-120V/50-60Hz, 700VA (CNME010717)
AC220V- 230V/50Hz or AC110-120V/50-60Hz, 750VA (CNME010717B/CNME010717A)
Control ModePre-warm Mode, Manual Mode and Baby Mode are controlled by micro-computer
Baby Mode temperature control range34.5°C-37.5°C
Skin temperature display range5°C-65°C
Accuracy of skin temperature sensor±0.2°C
Temperature uniformity of mattress≤2°C
Angle of heater head0° , 30° , 60° , 90° for two ways
Inclination of the bassinet±10°
APGAR timeraudible and visible alarm when the device runs to 50"~1',4'50"~5',9'50"~10'
Failure alarmover temperature alarm, deviation alarm, sensor failure alarm, power failure alarm, setting alarm, checking alarm and so on.
The total irradiance Ebi on the effective surface area≥0.66mW/cm2 (CNME010717B/CNME010717A)
Average value of general bilirubin radiation on the mattress effective range≥0.58mW/cm2 ( CNME010717B/CNME010717A)
Bilirubin uniformity of radiating on the mattress effective range>0.4 (CNME010717B/CNME010717A)
Dominant wavelength of blue light420-470nm

Medical Device Infant Radiant Warmer for NewbornMedical Device Infant Radiant Warmer for NewbornMedical Device Infant Radiant Warmer for NewbornMedical Device Infant Radiant Warmer for Newborn