China Pharmaceutical Laboratory Glove Integrity Tester Automatic Isolatorisolator, Find details about China Medical Isolator, Laboratory Isolator from Pharmaceutical Laboratory Glove Integrity Tester Automatic Isolatorisolator
1. Portable structure design, easy to use;
2. All stainless steel case, easy to clean, suitable for use in clean areas;
3. Tilted control panel for easy viewing and operation;
4. The test gas is external compressed air and nitrogen;
5. Provide result printing function to save records;
6. Controlled by SIEMENS programmable controller (PLC) and touch screen, with friendly human-computer interaction interface;
7. Leak test can be performed on the cuff-gloves before or after use of the isolator;
Input pressure | 0.2MPa~0.7MPa |
Inflation pressure | 500 ~ 1000Pa |
Test time | 5~8min |
Pressure display resolution | 0.1Pa |
Detection accuracy (minimum aperture) | 500μm |
Instrument weight | 19kg |
Mainframe dimensions(L*W*H) | 540mm×235mm×330mm |
Test port specification | 255mm (the rest of the specifications can be customized) |
Printing function | Pin mini printer |
Storage data | 20 groups |
Power supply | 220V, 50Hz |
Power | 100W |