China Smu Instrument Schottky, Rectifier Diode, IV Test, on-Voltage Test, Leakage Current Test Source Measurement Unit, Find details about China Smu Instrument, IV Test from Smu Instrument Schottky, Rectifier Diode, IV Test, on-Voltage Test, Leakage Current Test Source Measurement Unit
Max output power | 30w, four quadrant source or sink operation |
Source/sink limits | ±30V (≤1A), ±300V (≤100mA), |
±1.05A (≤30V), ±105mA (≤300V) | |
Overrange | 105% of range, source and measure |
On-load capacitance | <22nF |
Noise (20MHz) | 2mV RMS, <20mV Vp-p |
Voltage guard | Output impedance 1 kΩ, Deflection-Voltage<80mV |
Max sampling rate | 1000 sample/s |
Trigger | I/O |
Communication Interface | RS-232,GPIB |
Power supply | AC100V~240V 50/60Hz |
Temperature coefficient | 25±10 ºC |
Length,Width and Height | 425mm L x 255mm W x 106mm H |
Weight | 5KG |
Warranty | 1 year |
Voltage | Source | Measure | ||
Range | Programming resolution | Accuracy±(% rdg.+ V) | Programming resolution | Accuracy±(% rdg.+ A) |
300mV | 30uV | 0.1%±300uV | 30uV | 0.1%±300uV |
3V | 300uV | 0.1%±500uV | 300uV | 0.1%±500uV |
30V | 3mV | 0.1%±3mV | 3mV | 0.1%±3mV |
300V | 30mV | 0.1%±30mV | 30mV | 0.1%±30mV |
Current | Source | Measure | ||
Range | Programming resolution | Accuracy±(% rdg.+ V) | Programming resolution | Accuracy±(% rdg.+ A) |
100nA | 10pA | 0.1%±0.5nA | 10pA | 0.1%±0.5nA |
1uA | 100pA | 0.1%±3nA | 100pA | 0.1%±3nA |
10uA | 1nA | 0.1%±5nA | 1nA | 0.1%±5nA |
100uA | 10nA | 0.1%±50nA | 10nA | 0.1%±50nA |
1mA | 100nA | 0.1%±300nA | 100nA | 0.1%±300nA |
10mA | 1uA | 0.1%±5uA | 1uA | 0.1%±5uA |
100mA | 10uA | 0.1%±20uA | 10uA | 0.1%±20uA |
1A | 100uA | 0.1%±2mA | 100uA | 0.1%±2mA |