CE Marked ICU Ventilator Cwh-3020b

China CE Marked ICU Ventilator Cwh-3020b, Find details about China ICU, Ventilator from CE Marked ICU Ventilator Cwh-3020b

Model NO.
CWH-3020B
Model NO.
CWH-3020B
 
  1. General requirements
1.1 Patient range: Support adult and pediatric;
1.2 Display: ≥12 inch high definition TFT colour screen (touchable), can display 3 types of waveform and loops in same screen;
1.3 Power: AC 100-242V, 50/60Hz, with back-up battery, can be used for more than 120 mins after fully charge;
1.4 Working type: Pneumatic driven,electronically controlled;
1.5 Pipeline input range: 0.28~0.6MPa;
1.6 Apply with Day/Night mode,Standby;
1.7 One-key lock for screen, avoid the mis-operation;

2. Ventilation mode and typical function
2.1 Standard ventilation mode: VCV, PCV, PSV, SIMV, APNEA (A/C), , SIMV/PSV,  CE Marked ICU Ventilator Cwh-3020bCE Marked ICU Ventilator Cwh-3020bCE Marked ICU Ventilator Cwh-3020bCE Marked ICU Ventilator Cwh-3020bCE Marked ICU Ventilator Cwh-3020b,Manual mode etc;
2.2 Apply with Invasive and Non-invasive function;
2.3 Apply with pressure and flow trigger function;
2.4 Apply with Synchronous nebulization and humidify function. High flow oxygen remedy function;

3. Ventilator parameter
3.1 Tidal volume range: 20-2000ml
3.2 Breathing rate: 1-100bpm (1-40 bpm under SIMV);
3.3 Inspiration time: 0.2-12s
3.4 Pressure control/Pressure support:0~80cmH2O
3.5 Inspiration plateau: OFF, 1~50%
3.6 SIGH: OFF, 1~5/100
3.7 Electronic PEEP: OFF,3~40 cmH2O
3.8 CPAP:4~20cmH2O
3.9 O2 concentration: 21%~100%, continuously adjustable
3.10 Inspiration trigger sensitivity: Pressure: OFF, -20~0cmH2O (based on PEEP)
                        Flow:0.5~30L/min

4. Monitoring parameter
4.1 Self-test when start-up the machine: include pressure, flow, O2 sensor etc
4.2 Parameter monitoring:
-Tidal volume monitoring: VT for inhalation and exhalation;
-Pressure monitoring: Ppeak, Pmean, pressure control level,pressure support level ,PEEP, Pplateau, Peak Inspiratory Negative pressure ;
-Breathing rate monitoring: Total breathing rate, breathing rate for spontaneous breath, breathing rate for mechanical ventilation;
-Volume monitoring: minute volume, minute volume for spontaneous breath, minute leakage volume;
-Flow monitoring: maximum inspiration flow, maximum expiration flow;
-Resistance monitoring: inspiration resistance, expiration resistance;
-Gas monitoring: O2 concentration;
-Special monitoring: Flow trigger depth, Pressure trigger depth, Dynamic Compliance, Static compliance, RVR(RSBI);


Waveform monitoring: P-T,F-T,V-T waveform;
          Loops monitoring: P-V, V-F,F-P loops;

Alarm: VT, MV , airway pressure, continuous pressure, breathing frequency, no connection of tubing, Apnea, O2 concentration,PEEP, O2 or Air failure, AC power failure, back-up battery low, machine fault, etc. Comprehensive 3 level alarm function include audible (adjustable), visual and flicking alarm.
 

 
5.Air compressor
5.1 Air compressor with same brand, main unit of compressor
5.2 Sustained flow: ≥45L/min
5.3 Export pressure: 0.4 Mpa