China Hdt Plastic Vicat Determination Machine Used ISO/ASTM Standard, Find details about China Thermal Deformation Test Machine, Thermal Deformation Testing Machine from Hdt Plastic Vicat Determination Machine Used ISO/ASTM Standard
Temperature control range | -300 ºC ambient temperature |
Heating rate | (120 ± 10) ºC / h (12 ± 1) ºC / 6min (50 ± 5) ºC / h (5 ± 0.5) ºC / 6min |
Temperature indication error | 0.1 ºC |
Temperature control accuracy | ± 0.5 ºC |
Maximum deformation indication error | ± 0.01mm, |
Deformation Measurement range | 0-1 .5 mm |
Sample holder Count | 3 |
Quality rods and pallet load | 68g |
Heating medium | methyl silicone oil (kinematic viscosity of 200 centistokes usually selected) or transformer oil |
Cooling | natural cooling more than 150, 150 or less water cooling or natural cooling. |
Heating power | 4kw |
Instrument Size | 540mm × 520mm × 970mm |
main machine | 1 set |
test frame | 3 |
thermal deformation pressure head | 3 |
vicat pressure needle | 3 |
displacement sensor | 3 |
temperature sensor | 1 |
computer with good brand | 1 set (512M DDR/80G SATA/no floppy drive /DVD/Integration card /integration video /antibiosis waterproof keyboard /photoelectricity antibiosis mouse // WIN XP HOME Chinese version,15 LCD) |
color printer with good brand | 1 set |
test soft based on the WIN98\2000\XP Chinese version | 1 set |
power line | 1 stick |
signal line | 1 stick |
weights | 1 set |
Name Specification | A | B | C | D | E | F | G | H | I | J | K | W |
Weight (g) | 1000 | 900 | 500 | 200 | 100 | 50 | 20 | 10 | 5 | 2 | 40 | 60(load stem) |
Piece | 12 | 3 | 3 | 6 | 3 | 3 | 6 | 3 | 3 | 6 | 3 | 3 |