HP-Chem2200s Biochemistry Analyzer - Touch Screen Style

China HP-Chem2200s Biochemistry Analyzer

Model NO.
HP-CHEM2200S
Colorimetric Pool
Titanium Alloy
Test Method
Absorbency, End-Poit, Kinetic, Two-Point, Standa
Display
7.5LCD Liquid Crystal Display
Volume
470*320*230nm
13)Tweight
9kg
Fuse
φ5×20 2A 2
2)Trelative Humidity
≤70%
Trademark
HEPO
Transport Package
Exporting Wooden Package
Specification
CE SGS ISO9001 ISO13485 FDA
Origin
China
HS Code
9027809900
Model NO.
HP-CHEM2200S
Colorimetric Pool
Titanium Alloy
Test Method
Absorbency, End-Poit, Kinetic, Two-Point, Standa
Display
7.5LCD Liquid Crystal Display
Volume
470*320*230nm
13)Tweight
9kg
Fuse
φ5×20 2A 2
2)Trelative Humidity
≤70%
Trademark
HEPO
Transport Package
Exporting Wooden Package
Specification
CE SGS ISO9001 ISO13485 FDA
Origin
China
HS Code
9027809900
1. INTRODUCTION HP-Chem2200s Biochemistry Analyzer - Touch Screen Style
The HP-CHEM2200S Biochemistry Analyzer (Following abbr instrument) is a high performance instrument which can be used in the hospital and other correlated places for biochemical tests. Its characteristic is simplicity of operation, high sensitivity and exact detection data and so on. This instrument safe classification belongs to B model of 1.It has liquid crystal display to realize the man-machine conversation of programming and detection, besides its setting and detecting operation are easy and straightforward.
This instrument is mainly used to do the detection and analyzer of the clinic biochemistry index in hospital and other correlated department.
Notice: We do not provide reagent, so you should buy them by yourselves. Be sure that the reagents you buy are biochemical reagents and enrolled in your country.
2. Features
  1. The instrument can do detection, program and enquiry as well as drawing quality control curve.
  2. The instrument can presorte at least 200 programs of detection item and at least 5000 detection results.
  3. The instrument can show and print calibrating curve, Kinetic reaction curve, and print quality control curve.
  4. The instrument has the function of monocrotic and dual wavelength detection;
  5. The instrument can detect by the means of absorbance, kinetics, terminal, two-point, standard curve, blank comparing pipe and factor. It can print single report in real time, and also can print general report.
  6. The instrument will auto-check first after it is switched on and it also can communicate.
  7. The instrument has half press dormancy function.
  8. RS232 Communication which can connected to PC,thus can manage,collect and printe the data..Measured values can be transmitted to PC for futhur processing
 
3. Specifications
  1. The accuracy of filter's central wavelength and its semibreadth should accord with the rule of Table 1.
       Table 1 Central wavelength accuracy and semibreadth            Unit: nm
Wavelength rangeCenter wavelength accuracySemibreadth
340-750±3≤12
 
Standard configuration: 340,405,492,510,546,578, 630;
  1. Stray light: detecting with standard solution in 340nm wavelength, its absorbance is more than 2.0.
  2. Linearity error: the linearity error of absorbance analysis should be less than 1.0%.
  3. Repeatability: the repeatability of absorbance analysis should be less than 0.005.
  4. Stability: in schedule time, its absorbance change is less than 0.01.
  5. Temperature accuracy: when the temperature of liquid that will be detected is 37ºC,30ºC and 25ºC, and the accuracy is ±0.2ºC, the undulant value should be less than 0.4ºC.
  6. Cross pollution rate: the cross pollution rate should be no more than 1.0%.
  7. Accuracy of prove: with quality control serum as the sample to be detected, to do the detecting items of Kinetic and Terminal method respectively, the detection result should accord with the requirement of quality control serum manual.
  8. Printer: outlay USB printer.
  9. Communication: RS232 serial port (The communicated computer must accord with the standard.);
4. Work condition
  1. Environmental temperature: 10ºC~30ºC.
  2. Relative humidity:≤70%
  3. Power supply: 220 V, 50 Hz.
  4. Input power:160VA.
  5. Warm-up time: 10min.
  6. Keep away from strongly electromagnetic field.
  7. Avoid direct irradiation by strong light.
  8. Avoid dust, shake and there must be no caustic gas such as hydrochloric acid in air;
  9. Power supply must be earthed perfectly.