China Medical Diagnosis Equipment Multi-Parameter Bedside Patient Monitor, Find details about China Patient Monitor, Portable Patient Monitor from Medical Diagnosis Equipment Multi-Parameter Bedside Patient Monitor
Features:
RESP:
Method: impedance method
RR measurement range: Adult: 7-120bpm Pediatric/Neonate: 7-150bpm
Resolution: 1 turns per minute
RESP Apnea Alarm: 10s-40s
Alarm: Audible and visual alarm
Sweep speed: 6.25\12.5\25mms
Gain selection: X1, X2, X4
ECG:
Lead Type: 5 Lead cable: RA; LA; RL; LL or R; L; N; F; C
3 lead cable: RA; LA; LL or R; L; F
Gain selection: X0.25, X0.5, X1, X2
Sweep Speed: 12.5, 25, 50mm/s
Heart Rate Range: Adult: 15-300bpm
Pediatric/Neonate: 15-300bpm
Resolution: 1 bpm
Protection: Withstand 4000VAC/50HZ voltage in isolation
Against electrosurgical interference and defibrillation
Accuracy: ±1% or ±2bpm(whichever is greater)
Mode: Filter MON Mode: 0.5HZ~40HZ
DIA Mode: 0.05HZ~130HZ
OPE Mode: 1HZ~20HZ
ST Mode: 0.2HZ~40HZ
NIBP:
Work mode: Manual/Automatic/Continuous
Auto measure time: Adjustable(1-480min)
Measurement Unit: mmHg/KPA selectable
Measurement Type: Systolic, Diastolic, Mean
Range of systolic pressure: Adult: 40-270mmHg
Pediatric: 40-200mmHg
Neonate: 40-135mmHg
Range of diastolic pressure: Adult: 10-215mmHg
Pediatric: 10-150mmHg
Neonate: 10-100mmHg
Range of mean pressure: Adult: 20-235mmHg
Pediatric: 20-165mmHg
Neonate: 20-110mmHg
Over-pressure protection: both Hardware and software over pressure protection
Accuracy: Less than ±5mmHg
PR from NIBP: 40-240BPM
Resolution: 1mmHg
Temperature(Dual Channel):
Measurement range: 0-50ºC(32 to 122F)
Resolution: 0.1ºC
Accuracy: ±1ºC(without probe)
Channel: 2 channels, T1, T2, TD(Temperature Difference)
SPO2:
Measurement & alarm range: 0~100%
Resolution: 1%
Accuracy: ±2%(70~100%, Ped, non-motion+)
±3%(70~100%, Neo, non-motion); 0-69% unspecified
Alarm range: 0~100%
Pulse rate: Range: 20~250bpm
Resolution: 1bpm
Accuracy: ±3bpm(non-motion) or ±5bpm(motion)
Alarm range: 20-250bpm