Patient Monitor, 5 Parameter Patient Monitors, Portable Patient Monitor

China Patient Monitor, 5 Parameter Patient Monitors, Portable Patient Monitor, Find details about China Cms6000 Patient Monitor, 5 Parameter Patient Monitors from Patient Monitor, 5 Parameter Patient Monitors, Portable Patient Monitor

Model NO.
CMS6000
Oxygen Saturation(SpO2)
The Percentage of Hbo2 in The Total Hb in The Bloo
Channel
Dual-Channel
Paper Speed
25mm/S, 50mm/S
Resolution for CO2
0.1 mm Hg 0 to 69 mm Hg
Trademark
Mecan
Transport Package
Export Standard Packaging for Air and Sea
Specification
230 x 133 x 220(mm)
Origin
China (Mainland) Guangdong
Model NO.
CMS6000
Oxygen Saturation(SpO2)
The Percentage of Hbo2 in The Total Hb in The Bloo
Channel
Dual-Channel
Paper Speed
25mm/S, 50mm/S
Resolution for CO2
0.1 mm Hg 0 to 69 mm Hg
Trademark
Mecan
Transport Package
Export Standard Packaging for Air and Sea
Specification
230 x 133 x 220(mm)
Origin
China (Mainland) Guangdong

CMS6000  Patient Monitor, 5 parameter patient monitors, portable patient monitor

Model:  MC-CMS6000  

Patient Monitor, 5 Parameter Patient Monitors, Portable Patient Monitor
 

Introduction:

CMS6000 features in high cost-performance, high safety and low power consumption. To ensure monitoring quality and reduce operation risk, it adopts full-isolation(floating), ECG defibrillation-proof protection, ECG anti-high frequency surgical unit and NIBP dual-overpressure protection.

Features:

u  Display mode: 8'' color TFT LCD with high-resolution.

u  Be applicable for adult, pediatric and neonatal for all-round monitoring.

u  Operation interface with Chinese and English. (Optional languages: French, German, Italian, Dutch,Turkish, etc.)

u  Multiple display modes are optional, such as big characters interface.

u  Waveform and data colour can be set optionally.

u  Storage of 480-hour trend data, and review of 40-second holographic waveform.

u  7-lead ECG displays in a screen, and ECG waveforms series display.

u  Anti-high frequency surgical unit, defibrillation-proof(requirement for special leads).

u  Adopt digital SpO2 technology, which has strong anti-interference and anti-weak filling capability.

u  Function of NIBP review, storage for up to 400 NIBP data.

u  Network: connecting with central station, other Bed observation and software updating.

u  Network connection mode: wireless and wired.

u  Built-in rechargeable battery for uninterrupted monitoring.

u  Connect with external moving bracket.

u  Built-in thermal printer.

 

 

Specification:

Classification

The type of protection against electroshock

Class I equipment, and internally powered equipment

EMC

Class A

The degree of protection against electroshock

Type CF applied part

The degree of protection against ingress of water

IPX0

Dimension and weight

Dimension

230 x 133 x 220(mm)

Net weight

2.6kg

Display

Color TFT LCD

8 inch display

Resolution: 800 × 600

Working performance

Waveform

ECGI, ECGII, PLETH, RESP, ETCO2, 2 IBP

Time for trend review

480-hour

User interface

Multilevel menu for selecting, and can connect with central monitoring station

ECG

Lead Mode

3-lead and 5-lead are optional

Lead Selection

I, II, III, aVR, aVF, aVL, V

Gain

×0.25; ×0.5; ×1; ×2

Scan Speed

25 mm/s, 50 mm/s

Heart Rate

Range: 15~300bpm

Accuracy: ±1% or ±1bpm, which is greater

Alarm Indication

Auditory, visual and characters

ST Segment Monitoring

Monitoring Range: -0.6mV~+0.8mV

ARR

thirteen kinds of arrhythmia

RESPIRATION

Measuring Method

Impedance

Measuring Range

Adult                 0~120rpm

Pediatric and Neonatal   0~150rpm

Measuring Accuracy

±2 rpm

Apnea Alarm

10 ~ 40 S

Scan Speed

6.25mm/s, 12.5mm/s, 25 mm/s

NIBP

Measuring method

Oscillometry

Measuring range

a)Systolic: 40mmHg~270mmHg(adult)

         40mmHg~200mmHg(pediatric)

         40mmHg~135mmHg(neonatal)

b)Diastolic: 10mmHg~215mmHg(adult)

         10mmHg~150mmHg(pediatric)

         10mmHg~100mmHg(neonatal)

Cuff inflation

<30s(10~270 mmHg, standard adult cuff)

Measuring Period

Average<40s

Measuring Interval in AUTO Mode

1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120, 240, 480 min.

Accuracy

±10% or ±8mmHg, which is greater

TEMP

Channel

dual-channel

Measuring Range

0~50(ºC)

Accuracy

±0.1ºC

Resolution

0.1ºC

SpO2

Oxygen Saturation(SpO2)

The percentage of HbO2 in the total Hb in the blood

Measuring range

0~100%

0%~69% unspecified

Accuracy

±2bpm or ±2%, whichever is greater

Resolution

1%

Upper and lower limit for alarm

0%~100%, error: ±2%

Pulse rate(PR)

0~250bpm

Upper and lower limit for PR alarm

0bpm~250bpm, error: ±2bpm

Printer (optional)

Channel

dual-channel

Recording Width

48mm

Paper Speed

25mm/s, 50mm/s

EtCO2(optional)

Method

Sidestream or Mainstream

Measuring Range for CO2

0~150mmHg

Resolution for CO2

0.1 mm Hg   0 to 69 mm Hg

0.25 mm Hg 70 to 150 mm Hg

Accuracy for CO2

0 - 40 mm Hg         ±2 mm Hg

41 - 70 mm Hg         ±5%

71 - 100 mm Hg       ±8%

101 - 150 mm Hg     ±10%

Respiration rate>80BPM ±12%

AwRR Range

2~150BPM

Apnea Alarm

Available

IBP(optional)

Channel

dual-channel

Label

ART PA CVP RAP LAP ICP P1 P2

Pressure range for sensor

-50~350 mm Hg

Resolution

1 mm Hg

Accuracy

±2% or 1mm Hg, which is greater

Environment requirements

Temperature

Working: 5~40(ºC)

Storage and transportation: -40~+55(ºC)

Humidity

Working: 30%~75%

Storage and transportation:≤95%(no condensation)

Power supply

DC 9V

220V, 50Hz

Pmax: 150VA