ASTM D92 Flash and Fire Point Apparatus

China ASTM D92 Flash and Fire Point Apparatus, Find details about China Flash and Fire Point Apparatus, Cleveland Open Cup Tester from ASTM D92 Flash and Fire Point Apparatus

Model NO.
GD-3536A
Ambient Temperatureambient Temperature
(10~40)ºC
Relative Humidity
≤80%
Maximum Power Consumption
400W
Product Name
ASTM D92 Flash and Fire Point Apparatus
Trademark
GOLD
Transport Package
Standard Wooden Case
Origin
Chongqing
HS Code
9031809090
Model NO.
GD-3536A
Ambient Temperatureambient Temperature
(10~40)ºC
Relative Humidity
≤80%
Maximum Power Consumption
400W
Product Name
ASTM D92 Flash and Fire Point Apparatus
Trademark
GOLD
Transport Package
Standard Wooden Case
Origin
Chongqing
HS Code
9031809090

GD-3536A ASTM D92 Flash and Fire Point Apparatus


Introduction:
GD-3536A ASTM D92 Flash and Fire Point Apparatus is designed and made as per GB/T3536-2008 Petroleum Products - Determination of Flash and Fire Points - Cleveland open cup method and ASTM D 92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester. Cleveland Open Cup Tester is applicable to all petroleum products with flash points above 79ºC and below 400ºC except fuel oils.



I. Main technical features
1. Cleveland Open Cup Flash Point Tester adopts LCD screen to display and full English man-machine interface. It can preset expected flash point temperature, sample mark number, barometric pressure and other parameters.It has menu prompt and input guide.
2. COC Flash Point Apparatus adopts simulation tracking to display the function curve of temperature rising and test time. It has the functions of English mis-operation prompt,test date,test time and other parameters.
3. Automatically calibrate the effect of atmospheric pressure to test and calculate the corrected value.
4. Differential coefficient detection.Automatically correct the system deviation.
5. Automatically open the lid,detect the flash point and print test data.The test arm automatically rise up and lower down.
6. Reasonable design, safe and convenient to operate.


II. Main technical specifications
Power supplyAC220(-10%~+5%)V, 50Hz
Flash point determinationRange:Ambient to 400ºC
Repeatability:≤8ºC
Reproducibility:≤17ºC
Accuracy:0.1ºC
Heating rateCorrespond with GB/T3536 and ASTM D92
Ignition modeElectric ignition. 
Gas flame diameter is 3.2mm~4.8mm
Ambient temperature(10~40)ºC
Relative humidity≤80%
Maximum power consumption400W
Dimension410*360*310mm