Biobase China Counter Hematology Analyzer Cell Counter Hematology Analyzer, Find details about China Hospital Machine, Auto Hematology Analyzer from Biobase China Counter Hematology Analyzer Cell Counter Hematology Analyzer
Product Description BK-6190 Auto Hematology Analyzer
. 3 part diff of WBC. . 20 parameters+3 histograms. . 60 tests per hour..touch screen
Features of Touch Screen Constant-Temperature Incubator BJPX-H54BK(D/G)
Features:
.Throughput: 60T/H
.8 inch touch screen
.20 parameters + 3 histograms
.3 counting modes
.100,000 sample results
.Support LIS and external printer
Model | BK-6190 | ||||
Throughput | 60Tests/hour | ||||
Measuring Principle | Impedance for cell counting Cyanide-free method for HGB | ||||
Aperture Diameter | WBC chamber:100μm,RBC/PLT chamber:70μm | ||||
Parameters | WBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR | ||||
Histograms | 3 histograms(includingWBC, RBC, PLT histogram) | ||||
Sample Volume | Normal:10μl; Predileted:20μl | ||||
Calibration | Manual & auto calibration | ||||
Counting Modes | Venous, capilary, prediluted | ||||
Printer | Built-in thermal printer,USB to external printer(optional) | ||||
Data Input | 8 inch touch screen, keyboard, mouse | ||||
Output | RS232 port, USB port, VGA port. LIS system avaliable | ||||
Storage | 100,000 results with histogram | ||||
Carry-over rate | Parameter | CV | |||
WBC | ≤ 2% | ||||
RBC | ≤ 1% | ||||
HGB | ≤ 2% | ||||
PLT | ≤ 2% | ||||
Background | Parameter | Background | |||
WBC | ≤ 0.2×10^9/L | ||||
RBC | ≤ 0.02×10^12/L | ||||
HGB | ≤ 1 g/L | ||||
HCT | ≤ 0.5% | ||||
PLT | ≤ 10×10^9/L | ||||
Precision
| Parameter | CV | Measurement range | ||
WBC | ≤ 2.0% | (4.0-15.0×10^9)/L | |||
RBC | ≤ 1.9% | (3.5-6.0×10^12)/L | |||
HGB | ≤ 1.9% | (110.0-180.0g)/L | |||
MCV | ≤ 0.4% | (80.0-110.0)fL | |||
PLT | ≤ 4.0% | (100.0-500.0×10^9)/L | |||
Linearity | Parameter | Measurement range | CV | ||
WBC | (0-5.9×10^9)/L (6.0-99.9×10^9)/L | (±0.3×10^9)/L ±5% | |||
RBC | (0-0.99×10^12)/L (1.0-9.99×10^12)/L | (±0.05×10^12)/L ±5% | |||
HGB | (0-99.0g)/L (100.0-300.0g)/L | (±2.0g)/L ±2% | |||
PLT | (0-99×10^9)/L 100-999×10^9)/L | (±8.0×10^9)/L ±10% | |||
Power supply | AC 220±10%,60/50Hz, 110±10%,60Hz | ||||
Package Size | 580*470*590mm (W*D*H) | ReagentPackage Size | 450*410*310mm (W*D*H) | ||
Gross Weight | 27kg | ReagentGross Weight | 23kg |