Biobase Clinical Cell Counter Blood Testing 3-Part Automatic Hematology Analyzer

China Biobase Clinical Cell Counter Blood Testing 3-Part Automatic Hematology Analyzer, Find details about China Hematology Analyzer, Hematology Analyser from Biobase Clinical Cell Counter Blood Testing 3-Part Automatic Hematology Analyzer

Model NO.
BK-6190
Aperture Diameter
Wbc Chamber:100μm, Rbc/Plt Chamber:70μm
Histograms
3 Histograms(Includingwbc, Rbc, Plt Histogram)
Sample Volume
Normal:10μl; Predileted:20μl
Calibration
Manual & Auto Calibration
Counting Modes
Venous, Capilary, Prediluted
Printer
Built-in Thermal Printer, USB to External Printer
Data Input
8 Inch Touch Screen, Keyboard, Mouse
Output
RS232 Port, USB Port, VGA Port. Lis System Avaliab
Storage
100, 000 Results with Histogram
Power Supply
AC 220±10%, 60/50Hz, 110±10%, 60Hz
Trademark
BIOBASE
Transport Package
Wooden Box
Specification
Blood testing machine
Origin
China
Model NO.
BK-6190
Aperture Diameter
Wbc Chamber:100μm, Rbc/Plt Chamber:70μm
Histograms
3 Histograms(Includingwbc, Rbc, Plt Histogram)
Sample Volume
Normal:10μl; Predileted:20μl
Calibration
Manual & Auto Calibration
Counting Modes
Venous, Capilary, Prediluted
Printer
Built-in Thermal Printer, USB to External Printer
Data Input
8 Inch Touch Screen, Keyboard, Mouse
Output
RS232 Port, USB Port, VGA Port. Lis System Avaliab
Storage
100, 000 Results with Histogram
Power Supply
AC 220±10%, 60/50Hz, 110±10%, 60Hz
Trademark
BIOBASE
Transport Package
Wooden Box
Specification
Blood testing machine
Origin
China
BIOBASE BK-6190 Auto Hematology Analyzer
Product Description
Biobase Clinical Cell Counter Blood Testing 3-Part Automatic Hematology Analyzer
Features:
1.Throughput: 60T/H
2.8 inch touch screen
3.20 parameters + 3 histograms
4.3 counting modes
5.100,000 sample results
6.Support LIS and external printer

 

Product Parameters
Model
BK-6190
Throughput
60Tests/hour
Measuring Principle
Impedance for cell counting Cyanide-free method for HGB
Aperture Diameter
WBC chamber:100μm,RBC/PLT chamber:70μm
Parameters
WBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR
Histograms
3 histograms(includingWBC, RBC, PLT histogram)
Sample Volume
Normal:10μl; Predileted:20μl
Calibration
Manual & auto calibration
Counting Modes
Venous, capilary, prediluted
Printer
Built-in thermal printer,USB to external printer(optional)
Data Input
8 inch touch screen, keyboard, mouse
Output
RS232 port, USB port, VGA port. LIS system avaliable
Storage
100,000 results with histogram
Carry-over rate
Parameter
CV
WBC
≤ 2%
RBC
≤ 1%
HGB
≤ 2%
PLT
≤ 2%
Background
Parameter
Background
WBC
≤ 0.2×10^9/L
RBC
≤ 0.02×10^12/L
HGB
≤ 1 g/L
HCT
≤ 0.5%
PLT
≤ 10×10^9/L

Precision
Parameter
CV
Measurement range
WBC
≤ 2.0%
(4.0-15.0×10^9)/L
RBC
≤ 1.9%
(3.5-6.0×10^12)/L
HGB
≤ 1.9%
(110.0-180.0g)/L
MCV
≤ 0.4%
(80.0-110.0)fL
PLT
≤ 4.0%
(100.0-500.0×10^9)/L
Linearity
Parameter
Measurement range
CV
WBC
(0-5.9×10^9)/L
(6.0-99.9×10^9)/L
(±0.3×10^9)/L
±5%
RBC
(0-0.99×10^12)/L
(1.0-9.99×10^12)/L
(±0.05×10^12)/L
±5%
HGB
(0-99.0g)/L
(100.0-300.0g)/L
(±2.0g)/L
±2%
PLT
(0-99×10^9)/L
100-999×10^9)/L
(±8.0×10^9)/L
±10%
Power supply
AC 220±10%,60/50Hz, 110±10%,60Hz
Package Size
580*470*590mm (W*D*H)
ReagentPackage Size
450*410*310mm (W*D*H)
Gross Weight
27kg
ReagentGross Weight
23kg
Company Profile

Biobase Clinical Cell Counter Blood Testing 3-Part Automatic Hematology AnalyzerBiobase Clinical Cell Counter Blood Testing 3-Part Automatic Hematology Analyzer

Exhibition


Biobase Clinical Cell Counter Blood Testing 3-Part Automatic Hematology Analyzer

Customer feedback


Biobase Clinical Cell Counter Blood Testing 3-Part Automatic Hematology Analyzer

Packaging & Shipping


Biobase Clinical Cell Counter Blood Testing 3-Part Automatic Hematology Analyzer


Contact Me:Liya Jia