China ASTM D93 LCD Display Pensky Martens Closed Cup Flash Point Tester, Find details about China ASTM D93 Flash Point Apparatus, ASTM D93 Flash Point Test Apparatus from ASTM D93 LCD Display Pensky Martens Closed Cup Flash Point Tester
PMCC-B Pensky Martens Closed Cup Flash Point Apparatus is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens Closed Cup Method and ASTM D93 Standard Test Method for Flash Point by Pensky-Martens Closed Cup Tester. It is used to determine the closed cup flash point of the petroleum products.
I. Main technical features
1. The PMCC-B Pensky Martens Closed Cup Flash Point Apparatus adopts single chip microcomputer control technique.LCD screen displays. English man-machine interaction interface. The heating power is continuously step-less adjustable. The rate of heating up is exact.Easy setting for parameters.
2. The LCD screen has prompting menu,prompting input operation interface. It shows the setted parameters, real-time temperatures and other test parameters. Press the "record"button when the flash point appears. The screen will show and reserve the flash point.
3. The structure is newly design and small.Stainless steel table-board looks clean and beautiful. It can be used as long as equipped with gas or other civilian fuels.
4.PMCC-A Pensky Martens Closed Cup Flash Point Apparatus is easy to use and test results exact.
II. Main technical specifications
1 | Power supply | AC (220±10%)V, 50Hz |
2 | Total power consumption | ≤650W |
3 | Heating device | (1)The furnace body is made of silicon carbide (2)Heating power is adjustable from (0-600) w. |
4 | Heating rate | Procedure A: (5~6)ºC/min, Procedure B:(1~1.5)ºC/min |
5 | Stirring device | (1)Stirring motor: BYGH101 stepping motor (2) Driving mode: flexible shaft (3) Shaft size: 8mm×40mm |
6 | Stirring rate | Procedure A: (90~120)RPM, Procedure B: (250±10)RPM |
7 | Oil cup | (1) Inner diameter: 50.7mm~50.8mm. (2) Depth: 55.7mm~56.0mm (3) The scribed line depth: 33.9mm~34.3mm. (4) Volume :about 70ml |
8 | Igniting source | gas(or other civilian fuels,the same below) |
9 | Thermometers | Mercury-in-glass thermometer.Specifications are as below: (1)Scale -5ºC~110ºC, division 0.5ºC. (2)Scale 20ºC~150ºC, division 1ºC. (3)Scale 90ºC~370ºC,division 2ºC. |
10 | Working condition | Ambient temperature:(15~35)ºC; Relative humidity:≤ 85% |
11 | Dimension | 340mm×330mm×380mm |
12 | Net weight | 9KG |