China in Stock Biobase 2000 Fully Auto Elisa Processor Medical Laboratory Elisa, Find details about China Fully Automated Elisa Processor, Elisa Microplate Processor from in Stock Biobase 2000 Fully Auto Elisa Processor Medical Laboratory Elisa
Auto ELISA Processor BIOBASE 2000
Features:
1 robotic arm, 1 pipetting probe(10~1000μl).
2 unit 96 well microplates(independent incubating).
3 unit reader& washer (auto reading and washing).
Model | BIOBASE 2000 | |
Sample Unit | Sample Racks | 552 sample positions(tubes 13mm) |
Pipetting | 8~1000ul | |
Dispensing system | 2 aspirating-dispensing probe( X-Y-Z movement) | |
Reagent Unit | Reagent Position | 23 reagent racks, editable |
Pipetting Range | 8~1000ul, 1ul stepping | |
Pipetting Precision | (100ul )CV≤0.5% | |
Washing Unit | Dispensing Time | 4 minutes to whole 96 to microplates |
Washing Probes | Two line 8-nozzle manifolds | |
Wash Containers | Cleaning fluid(15L),waste water(25L),Buffer 1(5L), Buffer 2(2L), Buffer 3(2L), Buffer 4(2L), with liquid level-sensing, liquid shortage and full alarm | |
Waste Containers | 25L with waste full sensor | |
Washing Residual | <2μl | |
Reading Unit | Reading Channels | 8 independent photometric channels, mono anddichromatic reading |
Absorbance | 0.0~3.0 OD | |
Spectral Range | 400-700nm | |
Optical Filters | 2standard filters (450,630nm); 2 more on request (405nm/450nm/492nm/630nm) | |
Reading Accuracy | ±1% from 0.0~1.5 OD, ±2% from 1.5 to 3.0 OD | |
Data Management | Software System | Windows 7 |
LIS System | Bi-direction,support HL7 protocol | |
Results | Absorbance and results reviewable by software | |
Printer | External optional | |
Working Conditions | Power Supply | AC220V±10%,50/60Hz, 110V±10%,60Hz, 400W |
Temperature | 10~30ºC | |
Humidity | 30-80% | |
Size & Weight | External Size (W*D*H) | 1235*710*880mm |
Net Weight | 146kg | |
Package Size (W*D*H) | 1380*860*1435mm | |
Gross Weight | 220kg |