Oxygen Flowmeter Use for Patients

China Oxygen Flowmeter Use for Patients, Find details about China Oxygen Regulator, Oxygen Inhaler from Oxygen Flowmeter Use for Patients

Model NO.
DSW 905B905B, 906B
Material
Brass
Safety Valve
0.35±0.05MPa
Trademark
DSW
Transport Package
Carton
Specification
CE ISO
Origin
China
HS Code
9019200000
Model NO.
DSW 905B905B, 906B
Material
Brass
Safety Valve
0.35±0.05MPa
Trademark
DSW
Transport Package
Carton
Specification
CE ISO
Origin
China
HS Code
9019200000
Oxygen Flowmeter Use for Patients
Type
Brand NameModel NumberPlace of Origin Applicable GasInput PressureOutput PressureSafety valveThe Range of Flow
Breathing Apparatus DSWDSW 905B905B,906BDSW ChinaOxygen   15MPa0.2-0.3MPa  0.35±0.05MPa 1-15L/min
Connection Thread StructureWet the cupHumidity ConditioningPackaging Details DimensionG.WN.WMaterial
G5/8,CGA540 Spring-piston200ML 200ML10PCS/CTN 53*44*18cm 11.5Kg10Kg  brass


Medical Oxygen Intake Device For Medical Supply 
 

Outside thread G5/8,
Flow 0-10L/min or 0-15L/min
Output pressure 0.2-0.3MPa,
Input pressure 150bar
 
gas inlet types: 
1.gas -specific  inlet  probe conforming to BS ,AMERICAN, GERMAN, FRENCH 
2.1/8'' NPT female-threadde inlet 
3.Bullnose nipple with 3/8'' BSPNUT.   Other country-specific direct and remote-fitting configurations are 
available- contact the office. 
gas outlet types: 
1.body has female thread -use BMI 'universal'tubing nipple 
2.body has a 9/16'' UNFmale thread -use tubing nipple(oxygen) 
mounting: 
1.direct-probe fits into wall-mounted gas -specific socket(antirotation device maintains flowmeter upright) 
2.remote-kit available with 2m colour-coded flexible hose,and a gas-specific connector the flowmeter is mounted 
remotely from  the gas source on: 
a)wall/machine-mouted V-plate 
b)medical rail. 
NOTE:the rail-mounting kit utilises a "universal' rail clamp,allowing attachment to BMI,ohmeda,vickers"modura",
din rail section. 
operation and use 
2.service & maintence 
do dot disassemble when connected to a source of pressure 
caution 
repairs must only be undertaken by suitably trained personnel. 
the pre-use checks in section 2, and the leakage check in section 3.2 must be carried out after any disassembly 
3.tube assembly 
outer tube 
if the cover tube is cracked, withdraw the unit from service and renew: 
1.disconnect the flowmeter from the pressure source 
2.open the flow control valve to relieve any contained pressure. 
3.hold the unit vertically (to retain the flow tube and ball )unscrew the cover tube and lift off. 
calibrated tube and indicator ball 
4.lift off the flow tube , taking care not to let the ball roll out 
5.remove the tube connector adaptor and the cover tube o-ring 
6.fit a new cover tube o-ring at the base of the body. 
7.fit a new tube adaptor. Check orientation,as illustrated. 
8.fit the flow tube , and ball over the adaptor. 
9.position the tube with the scale towards the front of the unit. 
10.fit the outer tube,and screw clockwise into the flowmeter body. 
4. environmental conditions 
store/operate between 5 and 25 degree 
protect exposed o-rings from light