China Kdn Semi-Automatic Kjeldahl Distillation System, Find details about China Kjeldahl Distillation System, Kjeldahl Nitrogen Analyzer from Kdn Semi-Automatic Kjeldahl Distillation System
Name: Automatic Kjeldahl nitrogen analyzer
Model: KDN-04C/KDN-08C/KDN-20C
Note: KDN Series Automatic Kjeldahl nitrogen analyzer, the digestive furnace is not included in standard configuration
Characteristics
1. Adopt microcomputer controlling process
2. Automatically control distills, add water, control water level, and cut off the water supply
3. Various security protections: digestive system security device, steam generator water shortage alarm, water level detection fault alarm
4. Instrument shell adopts specially made plastic spray steel; working area adopts ABS anticorrosive board. Avoid chemical reagent corrosion and mechanical surface damage, acid-proof and alkali proof.
5. Once detected fault, the controlling system will power off automatically
6. Adopt tap water source, wide adaptability, and requirement for the test is low
Technical parameters
Measuring varieties: grain, feed, food, dairy products, drink, soil, medicine, precipitate, chemicals, and so on
Working mode: semi-automatic
Water-in mode: tap water and distilled water
Sample quantity: Solid: 0.20g-2.00g,
Semisolid: 2.00g-5.00g
Liquid: 10.00ml-25.00ml
Measuring range: 0.1mgN-200mgN (mg nitrogen)
Recovery rate: ≥99% (relative error, including digestive process)
Distillation speed: 5-15 min/sample (according to sample amount)
Cooling water consumption: digestion:5L/min
distillation:3L/min
Repetition rate: relative standard deviation <±1%
Power: AV 220V/50Hz, 1000W
Water supply: water pressure >1.5MPa
Water temperature<20°C
Shape size: 380*320*670mm
Weight: 20Kg
Model | KDN-04C | KDN-08C | KDN-20C |
function difference | Tap water source; with 4 holes digital digest furnace | Tap water source; with 8 holes digital digest furnace | Tap water source; with 20 holes digital digest furnace |