Best After Sales Medical Anesthesia Machine with Ventilator S6100A
10.4"TFT LCD touch screen dispalys the Ventilation parameters,Alarm information and Oscillogram.Alternate button for dual control.
5 mechanical flowmeters, instantly know the fresh gas flow to your patient
Integrated breathing circuit design, ensure easy operating and keep tidy
Gas source: O2,N2O,AIR
Ventilation mode: IPPV,A/C,SIMV,SIGH,MANUAL
Vt:0,20-1500ml
I:E :2:1-1:6
Oscillogram:P-T,F-T,P-V
Gas supply
- Gas source O2(For medical use)
N2O(For medical use) Air(For medical use)- Input pressure range 280 kPa~600 kPa
- Output pressure of regulator
(gas cylinder supply) less than 440 kPa- Safety valve 780kPa
- Negative pressure air source ≥75 L/min
Flow rate
- Adjust range of flow meter
O2 0.1 L/min~10 L/minN2O 0.1 L/min~10 L/minAIR 0.1 L/min~10 L/min- precision(20ºC 101.3kPa,supply pressure 440 kPa)
The precision from 10 % of full scale to full scale is ±10 % of indicated valueThe precision lower than 10 % of full scale is class 4.- Linkage control of O2 and N2O flow rate
Adjust O2 concentration to no less than 25 % according to proportion.- Flow rate of oxygen flush valve 25 L/min~ 75 L/min
- Output pressure for fresh valve 30 kPa ~ 45 kPa
Power supply
- Voltage 100-240 V ~
- Frequency 50/60 HZ
- Input power 80VA
- Fuse
--Main power T2AH 250V--Internal power T4AH 250V Ventilation mode
--IPPV--A/C --SIMV Ventilation parameters
Specification
- Frequency
--Adjusting range 1 bpm ~ 100 bpm(Under SIMV:1 bpm~ 20 bpmAll mode except SIMV: 4 bpm~100 bpm )--Allowable error ± 15 %- Inhalation time(I:E)(Tinsp)
--Adjusting range 2:1(1:0.5) ~ 1:6 (Ti 0.2s ~ 6.0s)--Allowable error ± 15 % or ± 0.1 s, whichever is the greater- Tidal volume(VT)
--Adjusting range 0,20 mL ~ 1500 mL - Minute ventilation(MV)
--Max MV ≥18 L/min- PEEP
--Adjusting range 0 kPa ~ 3kPa--Allowable error ± 15 % or ±(0.3 kPa+10%setting value),whichever is the greater- Pressure trigger(PTr)
--Adjusting range - 2.0 kPa ~ 0kPa(Under PEEP)--Allowable error ±(0.1 kPa+10 %setting value)- Flow trigger(FTr)
--Adjusting range OFF, 0.5 L/min ~ 30 L/min--Allowable error >3L/min: ±20%≤3L/min: ±0.6 L/min- Holding time(Inhalation platform)(TIP)
--Adjusting range 0 ~ 3s(Less than 50% of inhalation time)--Allowable error ± 15 % or± 0.1 s,whichever is the greater- Sigh
--Adjusting range 0, 1/100 ~ 5/100- Apnoea ventilation
--Adjusting range OFF,5 s ~ 60 s- Max set working pressure
--pressure limit range 2kPa ~ 10kPa--Allowable error ± 15 % or±1 kPa, whichever is the greater- Max pressure ≤12.5 kPa
explanation:
1) set pressure limit value to produce maximum working pressure( the upper limit of high airway pressure alarm).
2) do not use negative pressure at exhalation.
3) maximum limit pressure is pressure of the safety valve.
Driven gas
- Gas source O2(For medical use)
- Input pressure range 280 kPa~600 kPa
Fresh gascompensation
- Compensation range 0 ~ 15 L/min
- gas composition O2, N2O, AIR and anaesthetic
Ventilation parameter Monitoring
- The fllowing are monitoring parameters under the environment of body temperature and pressure-saturated.
- Minimum exhalation volume maybe monitored is 50 mL/min.
- Fig. display:
Parameter | Range | Resolving power | Accuracy |
Frequency(Freq) | 0 bpm ~ 100 bpm | 1 bpm | ±15 % |
Tidal volume(VT) | 0~ 2000 mL | 10 mL | >100 mL , ±20% ≤100 mL, ±0.02 L |
Minute ventilation(MV) | 0~100L/min | 0.1 L/min | >1L/min, ±20% ≤1L/min, ±0.5 L/min |
(Ppeak) | 0~100 cmH2O | 1 cmH2O | ±(4%full scale+4%actual reading) |
O2 concentration | 15 % ~ 100 % | 0.1 % | ±[ 2.5 %(v/v)+2.5 %(concentration level)] |
Compliance | 1~1000mL/ cmH2O | 1mL/ cmH2O | |
- Waveform display:
--Time- airway pressure(under all modes)--Time-flow(under all modes)--pressure volume loop(all modes)Remarks:The machine cannot record all the adjusting and monitoring values.