Model NO.
S6100A
I:E
2:1-1:6
Oxygen Concentration
15%~100%
Gas Source
Air, O2, N2o
Sigh
0 (off) 1/100~5/100
Ventilation Mode
Ippv, A/C, Simv, Sigh, Manual
Voltage
100~240V
OEM
Yes
Vt
0,20-1500ml
Trademark
Superstar
Transport Package
Wooden Cases
Specification
920*970*1380 mm
Origin
China
HS Code
9018907090
Model NO.
S6100A
I:E
2:1-1:6
Oxygen Concentration
15%~100%
Gas Source
Air, O2, N2o
Sigh
0 (off) 1/100~5/100
Ventilation Mode
Ippv, A/C, Simv, Sigh, Manual
Voltage
100~240V
OEM
Yes
Vt
0,20-1500ml
Trademark
Superstar
Transport Package
Wooden Cases
Specification
920*970*1380 mm
Origin
China
HS Code
9018907090
S6100A Best After Sales Medical Anesthesia Machine with CE
 
10.4"TFT LCD touch screen dispalys the Ventilation parameters,Alarm information and Oscillogram.Alternate button for dual control.
5 mechanical flowmeters, instantly know the fresh gas flow to your patient
Integrated breathing circuit design, ensure easy operating and keep tidy

Gas source: O2,N2O,AIR
Ventilation mode: IPPV,A/C,SIMV,SIGH,MANUAL
Vt:0,20-1500ml
I:E :2:1-1:6
Oscillogram:P-T,F-T,P-V
 

Gas supply

  1. Gas source                                           O2(For medical use)
                                         N2O(For medical use)
     Air(For medical use)
  1. Input pressure range                         280 kPa~600 kPa
  2. Output pressure of regulator
(gas cylinder supply)                   less than 440 kPa
  1. Safety valve                                          780kPa
  1. Negative pressure air source           ≥75 L/min
 

 Flow rate

  1.  Adjust range of flow meter
O                                            0.1 L/min~10 L/min
N2O                                        0.1 L/min~10 L/min
AIR                                         0.1 L/min~10 L/min
  1.  precision(20ºC 101.3kPa,supply pressure 440 kPa)
The precision from 10 % of full scale to full scale is ±10 % of indicated value
The precision lower than 10 % of full scale is class 4.
  1. Linkage control of Oand N2O flow rate
Adjust O2 concentration to no less than 25 % according to proportion.
  1. Flow rate of oxygen flush valve                  25 L/min~ 75 L/min 
  2. Output pressure for fresh valve             30 kPa ~ 45 kPa     

 Power supply

  1. Voltage                                        100-240 V ~
  2. Frequency                                   50/60 HZ
  3. Input power                                 80VA
  1. Fuse                                                      
--Main power                    T2AH 250V
--Internal power              T4AH 250V
 

Ventilation mode

--IPPV
--A/C                                                 
--SIMV
 

Ventilation parameters

Specification

  1. Frequency
--Adjusting range                   1 bpm ~ 100 bpm
(Under SIMV:1 bpm~ 20 bpm
All mode except SIMV: 4 bpm~100 bpm )
--Allowable error                           ± 15 %
  1. Inhalation time(I:E)(Tinsp) 
--Adjusting range                    2:1(1:0.5) ~ 1:6 (Ti 0.2s ~ 6.0s)
--Allowable error                       ± 15 % or ± 0.1 s, whichever is the greater
  1. Tidal volume(VT)
--Adjusting range                        0,20 mL ~ 1500 mL                         
  1. Minute ventilation(MV)                    
--Max MV                                      ≥18 L/min
  1. PEEP
--Adjusting range                        0 kPa ~ 3kPa
--Allowable error                     ± 15 % or ±(0.3 kPa+10%setting value),
whichever is the greater
  1. Pressure trigger(PTr)          
--Adjusting range                      - 2.0 kPa ~ 0kPa(Under PEEP)
--Allowable error                  ±(0.1 kPa+10 %setting value)
  1. Flow trigger(FTr)                        
--Adjusting range                            OFF, 0.5 L/min ~ 30 L/min
--Allowable error                                  
    >3L/min:                    ±20%
≤3L/min:                     ±0.6 L/min
  1. Holding time(Inhalation platform)(TIP)  
--Adjusting range                            0 ~ 3s(Less than 50% of inhalation time)
--Allowable error                             ± 15 % or± 0.1 s,whichever is the greater
  1. Sigh
     --Adjusting range                  0, 1/100 ~ 5/100
  1. Apnoea ventilation
--Adjusting range                   OFF,5 s ~ 60 s
  1. Max set working pressure                                    
--pressure limit range                       2kPa ~ 10kPa
--Allowable error                                  ± 15 % or±1 kPa, whichever is the greater
  1. Max pressure                                      12.5 kPa
explanation:
1) set pressure limit value to produce maximum working pressure( the upper limit of high airway pressure alarm).
2) do not use negative pressure at exhalation.
3) maximum limit pressure is pressure of the safety valve.

Driven gas

  1. Gas source                                          O2(For medical use)
  1. Input pressure range                         280 kPa~600 kPa

Fresh gascompensation

  1. Compensation range                            0 ~  15 L/min
  2. gas composition                       O2, N2O, AIR and anaesthetic
 

Ventilation parameter Monitoring

  1. The fllowing are monitoring parameters under the environment of body temperature and pressure-saturated. 
  2. Minimum exhalation volume maybe monitored is 50 mL/min.
  3. Fig.  display:
ParameterRangeResolving powerAccuracy
Frequency(Freq)0 bpm ~ 100 bpm1 bpm           ±15 %
Tidal volume(VT)0~ 2000 mL10 mL>100 mL ,  ±20%
≤100 mL,  ±0.02 L
Minute ventilation(MV) 0~100L/min0.1 L/min>1L/min,  ±20%
≤1L/min,  ±0.5 L/min
(Ppeak)0~100 cmH2O1 cmH2O  ±(4%full scale+4%actual reading)
O2 concentration15 % ~ 100 %0.1 %±[ 2.5 %(v/v)+2.5 %(concentration level)]
Compliance1~1000mL/ cmH2O1mL/ cmH2O 
  1. Waveform display:
--Time- airway pressure(under all modes)
--Time-flow(under all modes)
--pressure volume loop(all modes)
Remarks:The machine cannot record all the adjusting and monitoring values.