Asphalt Dynamic Viscosity Tester for Sale

China Asphalt Dynamic Viscosity Tester for Sale, Find details about China Dynamic Viscosity Tester, Dynamic Viscometer from Asphalt Dynamic Viscosity Tester for Sale

Model NO.
GD-0620A
Power
Electronic
Usage
Dynamic Viscosity Tester
Test Samples
4 Samples
Operation Interface
Touch-Type Colored LCD
Maximum Power Consumption
1800W
Timing Mode
4 Built-in Timer. Can Do Timing for 4 Capillary VI
Capillary Viscometer Washer
Gd-0620-2 Vacuum Capillary Viscometer Washer
Warranty
12 Months
Product Name
Asphalt Dynamic Viscosity Tester for Sale
Trademark
GOLD
Transport Package
Wooden Case
Specification
590mm*430mm*630mm
Origin
Chongqing, China (Mainland)
HS Code
9013809000
Model NO.
GD-0620A
Power
Electronic
Usage
Dynamic Viscosity Tester
Test Samples
4 Samples
Operation Interface
Touch-Type Colored LCD
Maximum Power Consumption
1800W
Timing Mode
4 Built-in Timer. Can Do Timing for 4 Capillary VI
Capillary Viscometer Washer
Gd-0620-2 Vacuum Capillary Viscometer Washer
Warranty
12 Months
Product Name
Asphalt Dynamic Viscosity Tester for Sale
Trademark
GOLD
Transport Package
Wooden Case
Specification
590mm*430mm*630mm
Origin
Chongqing, China (Mainland)
HS Code
9013809000
Asphalt Dynamic Viscosity Tester for Sale 

Introduction:

This instrument is made as per standards T 0620-2000 Dynamic Viscosity Test for Bitumen (Vacuum Capillary Method) in JTG E20-2011 Standard Test Methods for Bitumen and Bituminous Mixtures for Highway Engineering and SH/T0557-1998 Test method for Viscosity of Petroleum Asphalt (Capillary Method) . It is used to determine the dynamic viscosity of sticky petroleum asphalts by vacuum capillary viscometer. It is also suitable to determine other materials which viscosity range is 0.0036~20.000Pa.S.

Features:

1. It has a data base for coefficients of capillary viscometers. It can save 10 groups of coefficients at most. Operator can callout the data whenever need.

2. It adopts high accuracy pressure sensor. The vacuum degree can be kept at 300mmHg±0.5mmHg. It can be shown on the screen clearly.

3. The system will ignore the results which timing is less than 60s and enter next measurement step. The system will calculate the viscosity automatically for the results which timing is beyond 60s.

4. Operator can choose and set automatic calculation of average value and bias percentage for 2~4 parallel tests at will. The calculation can be done for several times to remove the unsatisfied results for operator. The test report can be printed. 

Specifications:

1. Temperature control mode: Point controlled by digital temperature controller. 

2. Temperature control arrange: 0.00ºC~100.00ºC

3. Temperature control accuracy: ±0.01ºC

4. Pressure range: 300mmHg±0.5mmHg

5. Timing mode: 4 built-in timer. Can do timing for 4 capillary viscometers

6. Timing range: 0.0s~99999.9s(27.7h); Bias ≤0.02%

7. Measurement range: About 18 Pa.s~580000Pa.s

8. Test samples: 4 samples

9. Operation interface: Touch-type colored LCD

10. Power supply: AC(220±10%)V , 50Hz

11. Ambient temperature: 5ºC~50ºC

12. Relative humidity: ≤85%

13. Overall dimension: 590mm×430mm×630mm

14. Net weight: 32.5kg(no water in bath)

15. Maximum power consumption: 1800W


 Optional part: