Biometer Lab Furniture Protect Stainless Steel Biological Biosafety Cabinet

China Biometer Lab Furniture Protect Stainless Steel Biological Biosafety Cabinet, Find details about China Lab Biosafety Cabinet, Biosafety Cabinet from Biometer Lab Furniture Protect Stainless Steel Biological Biosafety Cabinet

Model NO.
SWAQG-CY-BHC-1000IIA2
Front Suction Wind Speed
0.35m+-0.025m/S
Air Tightness
≤10-6m/S (Under 500PA)
Vibration in Half-Peak
≤5µm
Illuminance
≥300lx
Power Supply
AC Single Phase 220V / 50Hz
Maximum Power
600W
Weight
<200kg
Size of Working Area
1000*700*680mm
Outline Dimensions
1200*780*1950mm
Trademark
BIOMETER
Transport Package
Wooden Box
Specification
1200*780*1950mm
Origin
Shandong
HS Code
8414609011
Model NO.
SWAQG-CY-BHC-1000IIA2
Front Suction Wind Speed
0.35m+-0.025m/S
Air Tightness
≤10-6m/S (Under 500PA)
Vibration in Half-Peak
≤5µm
Illuminance
≥300lx
Power Supply
AC Single Phase 220V / 50Hz
Maximum Power
600W
Weight
<200kg
Size of Working Area
1000*700*680mm
Outline Dimensions
1200*780*1950mm
Trademark
BIOMETER
Transport Package
Wooden Box
Specification
1200*780*1950mm
Origin
Shandong
HS Code
8414609011

 

Product Description
Biometer Lab Furniture Protect Stainless Steel Biological Biosafety CabinetThe biosafety cabinet is a negative pressure safety equipment widely used in scientific research, teaching, clinical testing and production in the fields of microbiology, biomedicine, genetic recombination, animal experiments and biological products. It adopts the current advanced air purification technology and the negative pressure cabinet design, to achieve the protection of the environment, personnel and samples. It is the basic safety protection equipment in the laboratory biological safety primary barrier.
Product Features

1. The constant airflow is 30% externally exhausted and 70% internally circulated, with negative pressure vertical laminar flow.
2. Using imported motor, the safety cabinet itself can automatically adjust the wind speed forever.
3. The UV lamp can only work normally when the front window is closed properly.
4. When the opening height of the working window exceeds the specified value, an audible and visual alarm will be issued.
5. The power output socket in the work area has a waterproof protective cover and overload protection.
6. The front window of the safety cabinet can be opened manually and is easy to operate.
7. Clean working environment, smooth, seamless, no dead ends, can be easily and thoroughly disinfected.
8. The movable table top can facilitate the cleaning of the space under the working table.
9. The filter can be replaced regularly and can be easily taken out from the front.
10. The white inner wall is non-reflective, eliminating eye fatigue.
11. The control part is set outside the working area, which can be easily inspected and maintained without any risk.
12. Humanized settings, which can effectively protect the safety of the human body

Technical Parameters
Cleanliness 100 Class ≥0.5µm
 CFU  ≤0.5/vessel .hour 
 Front Suction Wind Speed  0.35m+-0.025m/s 
 Air Tightness   ≤10-6m/s (under 500Pa) 
 Noise  ≤62dB 
 Vibration in Half-peak  ≤5µm 
 Illuminance   ≥300LX 
 Power Supply  AC Single Phase 220V / 50Hz 
 Maximum Power  600W 
 Weight   <200kg 
 Specification and Quantity of High-efficient Filter  995*600*38mm*1 
 Specification and Quantity of Fluorescent Lamp / UV Lamp   20W*1 / 20W*1 
 Size of Working Area  1000*700*680mm 
 Outline Dimensions  1200*780*1950mm 
Related Products

Biometer Lab Furniture Protect Stainless Steel Biological Biosafety CabinetBiometer Lab Furniture Protect Stainless Steel Biological Biosafety Cabinet

Latest News

Biometer Lab Furniture Protect Stainless Steel Biological Biosafety Cabinet

Company Profile

Biometer Lab Furniture Protect Stainless Steel Biological Biosafety Cabinet

 

 

 

FAQ

1.Why choose BIOMETER?
We are a good manufactuer and reliable trading company in lab furniture and lab equipment,  Offer OEM & ODM service. 
2.How about the sample of our product?
We could send you the samples but the freight and the sample are charged.
3.How about Biometer payment term?
T/T & L/C &Western Union etc(40% deposit, the balance before shipment).
4.How about delivery time?
Within 7-15 workdays after receiving deposit.
5.What is the shipping port?
FOB to Qingdao Port,China(Also can ship according to customers' requests).
6.How about the package?
Bubble film + Cotton +Standard export wooden case.
7.How to inspect the goods?
Products will be checked by our QC staff frist, then our project manager.
Client can come and check by themselves or third partycheck is available.