Electrical parameters | |
---|---|
Accuracy class | 0.05%, 0.1% |
Power supply | Single phase AC 220V±10% or 110V±10%, 50/60Hz |
Communication port | USB, RS232, RS485, LAN |
AC Voltage output | |
Range(U1,U2,U3) | 50V, 100V, 200V, 400V, 600V |
Adjustment range | (0 - 120)% RG |
Adjustment resolution | 0.01% RG, 0.1% RG, 1% RG, 10% RG |
Stability | 0.01% /1min |
Distortion | ≤0.2% (non-capacitive load) |
Max. output load | 25VA for each phase |
Accuracy | 0.05% RG |
AC Current output | |
Range(I1,I2,I3) | 0.5A, 1A, 2.5A, 5A, 10A, 20A |
Adjustment range | (0 - 120)% RG |
Adjustment resolution | 0.01% RG, 0.1% RG, 1% RG, 10% RG |
Stability | 0.01% /1min |
Distortion | ≤0.2% (non-capacitive load) |
Max. output load | 25VA for each phase |
Accuracy | 0.05% RG |
AC Power output | |
Active output stability | 0.01%RG/1min |
Reactive output stability | 0.02%RG/1min |
Active accuracy | 0.05% RG |
Reactive accuracy | 0.1% RG |
Frequency output | |
Adjustment range | 45-65Hz |
Adjustment resolution | 1Hz, 0.1Hz, 0.01Hz and 0.001Hz |
Resolution | 0.001Hz |
Accuracy | 0.002Hz |
Power factor output | |
Adjustment range | -1 to 0 to +1 |
Adjustment resolution | 0.0001 |
Resolution | 0.0005 |
Phase output | |
Adjustment range | 0°-359.999° |
Adjustment resolution | 10°, 1°, 0.1°, 0.01° |
Resolution | 0.001° |
Accuracy | 0.05° |
Harmonic configuration | |
Times | 2 to 31 |
Content | 0-40% |
Phase | 0°-359.999° |
Configuration error | (10% RD + 0.1%), RD refers to the configuration value of harmonic contents |
DC Voltage output | |
Range | 75mV, 75 V, 150 V, 300 V, 500V, 1000 V |
Adjustment range | (0-120)% RG |
Adjustment resolution | 0.01% RG, 0.1% RG, 1% RG, 10% RG |
Stability | 0.01% RG / 1 min |
Distorting | ≤0.2% (non-capacitive load) |
Output load | 25VA |
Accuracy | 0.05% RG |
Ripple contents | ≤1% |
DC Current output | |
Range | 0.5 A,1A, 2.5 A, 5 A, 10A, 20 A |
Adjustment range | (0-120)% RG |
Adjustment resolution | 0.01% RG, 0.1% RG, 1% RG, 10% RG |
Stability | 0.01% RG / 1min |
Distortion | ≤ 0.2% (non-capacitive load) |
Output load | 25VA |
Accuracy | 0.05% RG |
Ripple contents | ≤1% |
Energy Error | |
Active error | 0.05% RG |
Reactive error | 0.1% RG |
DC Input Voltage Measurement | |
Range | 0 to ±20V |
Measurement range | (0-120)% RG |
Accuracy | 0.01% RG |
Resolution | 0.001% RG |
DC Input Current Measurement | |
Range | 0-20mA |
Measurement range | (0-120)% RG |
Accuracy | 0.01% RG |
Resolution | 0.001% RG |
Mechanical parameters | |
Dimensions (W×H×D) (mm) | 460x430x185 |
Weight (kg) | 20 |
Environmental conditions | |
Working temperature | 0°C to 40°C |
Relative humidity | ≤85% |